8 Troubleshooting
8.7 Troubleshooting: Other components
Problem
Liquid is leaking from
the external air sensor
The internal tempera-
ture of the air sensor is
too high
I/O-box
Problem
The internal tempera-
ture of the I/O-box is
too high
The I/O-box is not de-
tected by the system
The system does not
detect Digital signals
correctly
Digital In always "0",
regardless of input sig-
nals
410
Possible cause and action
One of the connections is not tight enough
Tighten the connectors.
Hot surroundings
Decrease the room temperature. Maximum operating
temperature is 35°C.
Hardware error
Switch off the instrument and wait until the tempera-
ture has decreased. Restart the instrument. If this error
is recurrent, generate a System error report and con-
tact Service.
Possible cause and action
Hot surroundings
Decrease the room temperature. Maximum operating
temperature is 35 C.
Hardware error
Wrong Node ID
Make sure that the Node ID is (00) for the primary box,
and (01) for a potential secondary box.
Cables incorrectly connected
Make sure that the cables are connected correctly,
earth to earth etc.
Digital Out connections switched
Check that the digital Out cables are correctly connect-
ed, earth to earth etc.
ÄKTA avant User Manual 29035184 AE