Possible causes of a failed test
The table below describes possible causes of a failed test. When possible sources of error
have been checked and taken care of, run the test once again.
Cause
Disturbances caused by
air trapped in Quater-
nary valve or any of the
System pumps
Disturbances caused by
damaged pump piston
seals.
Unstable or incorrect UV
signal, or drifting base
line - faulty UV monitor
Wrong mixer chamber
size or faulty mixer
Faulty Quaternary valve
Faulty System pumps
Incorrect preparation of
buffer and tubing
ÄKTA avant User Manual 29035184 AE
Action
Make sure to prime all Q inlets and purge the System
pumps before method start, see Section 5.4.3 Prime Q in-
lets, on page 178.
Replace piston seals. See ÄKTA avant Operating
Instructions, Maintenance chapter.
See UV monitor U9-M and UV detector unit, on page 363
Change mixer chamber or replace the Mixer. See Sec-
tion 7.8.2 Replace the Mixer, on page 338.
See Quaternary valve, on page 383
See Section 8.6 Troubleshooting: Pumps, on page 397
Make sure that the system was correctly prepared, see
Prepare the Q valve test, on page 253.