Test And Calibration; Troubleshooting; The Display Doesn't Light Up; Service" Appears On The Status Display Line - GE OptiSonde User Manual

General eastern chilled mirror hygrometer
Table of Contents

Advertisement

Test and Calibration

Troubleshooting

The Display Doesn't Light
Up
"Service" Appears on the
STATUS Display Line
Maintenance
The procedures in this section effectively test and/or calibrate the
following aspects of the OptiSonde:
Startup and power supply voltage
Normal sensor operation
Front panel display
Digital and analog outputs.
The unit has been completely tested and calibrated at the factory, and
is ready to plug in and operate. As shipped, it meets all of our
published specifications.
When ordered as a complete hygrometer system with a chilled mirror
sensor and cable, it is verified at a number of points against a dew
point system that has been certified by the U.S. National Institute of
Standards and Technology (NIST). A Certificate of Compliance is
supplied with the unit to indicate traceability.
1. On the benchtop model, check the POWER switch on the rear
panel. Make sure it's ON.
2. Check the line cord. Make sure both ends are plugged in and that it
is plugged into a proper source of AC voltage.
3. Check the power supply. Make sure it is connected and has the
proper output voltage.
4. Check the fuse. Make sure the proper fuse size is installed. Make
sure the fuse is not open.
The text "Service Req" displayed on the STATUS line means service
is required. The most frequent problem requiring service is that the
mirror surface is contaminated and should be cleaned.
Clean and balance the sensor mirror (refer to Minor Maintenance of
Sensor Optics on page 5-1). Next, run the instrument through a
PACER cycle. If, at the end of the cycle, the Service Req status
appears again, repeat the cleaning and balancing procedure, or contact
the factory.
October 2007
5-7