6 Performance tests

6.4 Fraction collector F9-R test

6.4

Fraction collector F9-R test

Method description
The Fraction collector F9-R test checks the functionality of Fraction collector F9-R.
The method run takes approximately 3 minutes.
Note:
Required configuration
A correctly installed Outlet valve, and a correctly installed Fraction collector F9-R are
required to run the test.
Required material
The following materials are required:
Prepare the Fraction collector
F9-R test
Follow the instructions to prepare the system before method start.
Step
1
2
3
4
5
6
250
The fraction collector test is evaluated manually and no report is generated.
Distilled water
Syringe, 25 to 30 ml
7 tubes, 2 ml or larger, for collecting the fractions.
Action
Make sure that the Fraction collector F9-R is connected to port Out 10 on
Outlet valve V9-O or V9H-O
Direct outlet tubing W to a waste container.
Place inlet tubing A1 into distilled water.
Prime inlet A1 and purge System pump A. See Section 5.4.1 Prime buffer inlets
and purge system pumps, on page 166.
Place 7 tubes in the Fraction collector, in positions 1 to 7.
In the System Control module, on the System menu, click Settings
Result: The System Settings dialog box opens.
ÄKTA avant User Manual 29035184 AE