5 Operation
5.8 Fractionation
5.8.3 Fractionation overview
Delay volume
The delay volume settings are used to make sure that the fractions collected during
fractionation, using the outlet valve or the fraction collector, correspond to the fractions
indicated in the chromatogram. The delay volume is the volume between the UV monitor,
and the fraction collector or outlet that is used, see the following illustration.
The time when a peak is detected
by the UV/Vis detector is called T
As the delay volume is affected by the length and diameter of the tubing, it should be
set according to the tubing and modules used, see Section 9.12 Delay volumes, on
page 505.
220
UV/Vis
pH
Conductivity
Delay volume
0
Valve
Fraction collector
The peak reaches the fraction
collector at time T