8 Troubleshooting
8.7 Troubleshooting: Other components
Problem
Weak light
The internal tempera-
ture of the instrument
display is too high
The cursor on the dis-
play touch screen does
not move
The cursor on the dis-
play touch screen does
not follow the finger
The display is black
412
Possible cause and action
Display is worn out
Generate a System error report and contact Service.
The main board for the display is overheated
Switch off the instrument and wait until the tempera-
ture has decreased. If this error is recurrent, generate
a System error report and contact Service.
Touch screen is damaged
Generate a System error report and contact Service.
Touch screen is not calibrated
Calibrate touch screen. See Section 2.5 Instrument
display, on page 37.
No power
Make sure that the power cord is connected.
Make sure that the instrument is on.
Display is broken
Generate a System error report and contact Service.
ÄKTA avant User Manual 29035184 AE