GE AKTA avant User Manual page 344

Hide thumbs Also See for AKTA avant:
Table of Contents

Advertisement

7 Maintenance
7.8 Replacement procedures
7.8.4 Replace the UV monitor U9-M flow cell
Step
6
7
8
9
Clean the optical fiber
connectors
Follow the instruction to clean the optical fiber connectors.
Step
1
2
344
Action
Pull off the black protective caps from the new UV flow cell, and connect the
new UV flow cell to the detector.
Connect the detector, with the new flow cell connected, to the monochro-
mator. Pull the latch upwards to fasten the detector.
Connect the tubing to the new flow cell.
Switch on the instrument.
Result: The flow cell path length is automatically recognized by the monitor
when a new flow cell is connected.
WARNING
Hazardous substances. When using hazardous chemicals, take
all suitable protective measures, such as wearing protective
clothing, glasses and gloves resistant to the substances used. Fol-
low local and/or national regulations for safe operation and
maintenance of the product.
Action
Wipe the optical fiber connectors with isopropanol on lens paper.
Wipe the optical fiber connectors dry with lens paper.
ÄKTA avant User Manual 29035184 AE

Advertisement

Table of Contents
loading

Table of Contents