4.5

Extra Outlet valves

Introduction
When using the standard configuration with one Outlet valve, 10 outlet positions are
available. To increase the number of outlets, one or two extra Outlet valves can be con-
nected, adding up to a total of 21 or 32 outlet positions. This optional configuration is
convenient when collecting a number of large fractions outside the fraction collector.
Locations and illustrations
It is recommended to install the extra outlet valves in the middle and in the lower position.
The illustration below shows the recommended positions.
Flow path
The tubing from the pH monitor is connected to the In port of the Outlet valve (V9-O or
V9H-O). The Out 10 port of the Outlet valve (V9-O or V9H-O) is connected to the In port
of the second Outlet valve (V9-O2 or V9H-O2). The Out 21 port of the second Outlet valve
is connected to the In port of the third Outlet valve.
The following illustration shows the flow path with two extra Outlet valves installed on
ÄKTA avant 25.
ÄKTA avant User Manual 29035184 AE
4 Optional instrument modules

4.5 Extra Outlet valves

117