Wash settings
The settings for system wash are accessible as parameters to the Advanced:System
and pump wash settings and Advanced:Loop wash settings instructions.
Parameter for System
and pump wash settings
System flow rate
System pump wash vol-
ume
Sample flow rate
Sample pump wash vol-
ume
Parameter for loop wash
settings
Loop wash flow rate
Loop wash high alarm
ÄKTA avant User Manual 29035184 AE
Description
Sets the flow rate used for the Pumps and pressures:Sys-
tem wash instruction.
Note:
The volume for system wash is set in the Pumps and
pressures:System wash instruction.
The flow rate should not exceed 10 ml/min if narrow
inlet tubing (i.d. 0.75 mm) is used.
Adjust the flow rate during the system wash so that
the system pressure does not exceed 2 MPa.
Sets the wash volume used during system pump
washes.
Sets the flow rate used during sample pump washes.
Note:
The flow rate should not exceed 10 ml/min if narrow inlet
tubing (i.d. 0.75 mm) is used.
Sets the wash volume used during sample pump
washes.
Description
Sets the flow rate used during Loop wash.
Note:
The volume for loop wash is set in the Pumps and
pressures:Loop wash instruction.
The flow rate should not exceed 10 ml/min if narrow
inlet tubing (i.d. 0.75 mm) is used.
Sets the maximum pressure value during Loop wash.
If the system pressure exceeds this value an alarm is
issued and the method is paused.