4 Optional instrument modules
4.12 I/O-box

4.12.4 Connect external equipment to the I/O-box

4.12.4
Connect external equipment to the I/O-box
Requirements on connected
equipment
The physical requirements for the connected equipment is described in the following
tables. All connected equipment must have a common grounding.
Analog input
Parameter
Channels
Range
Input impedance
Accuracy
Analog output
Parameter
Channels
Range
Input impedance
Accuracy
Digital input
Parameter
Channels
Compatibility
Digital output
Parameter
Channels
Compatibility
150
Value
2
±2000 mV
1 MΩ
±(0.1% + 0.2 mV)
Value
2
±1000 mV
100 kΩ
±(0.3% + 1 mV)
Value
4
TTL, open/closed circuit
Value
4
Open/closed circuit
ÄKTA avant User Manual 29035184 AE