9 Reference information
9.7 System settings

9.7.9 System settings - Advanced

9.7.9
System settings - Advanced
The following table describes the advanced system settings available for ÄKTA avant.
Instruction name
Constant pressure flow
parameters
Pressure control parame-
ters
Max flow during valve
turn
Method progressing flow
Sample pump setting
Instrument display
472
Description
Sets the values for the P and I factors needed to keep
a constant pressure by varying the flow rate. The signal
used for pressure control is set in the instruction Ad-
vanced:Constant pressure flow.
By using Pressure control the method can be run with
the set flow rate without the risk of method stop due to
pressure alarm. Pressure control is enabled in the instruc-
tion Pumps and Pressures:System flow or Pumps and
PressuresSample flow. Pressure control parameters
provides the P and I factors used in the regulator and
can be adjusted for different columns.
Sets the maximum flow rate used during the turning of
the injection and outlet valve in order to avoid high
pressure alarms.
Sets which flow (Automatic, System flow or Sample
flow) to use to calculate the progress of the method. In
automatic mode, the position of the injection valve de-
termines if the system flow or the sample flow is used.
Enables or disables sample pump flow while the injection
valve is in manual load position.
Locks or unlocks the Pause and Continue buttons on
the instrument display.