Problem
The internal tempera-
ture of the fraction col-
lector is too high
The fraction collector
failed to detect a drop
and has changed tube
without DropSync.
The tube sensor has
not detected a new
tube.
The delay queue is full
or there is a tube
change overload.
ÄKTA avant User Manual 29035184 AE
Possible cause and action
Hot surroundings
Decrease the room temperature. Maximum operating
temperature is 35°C.
Hardware error
Switch off the instrument and wait until the tempera-
ture has decreased. Restart the instrument. If this error
is recurrent, generate a System error report and con-
tact service.
To high flow rate for DropSync
Decrease the flow rate or disable DropSync.
Air in the flow path
Check the flow path for air. Fill system and purge
pumps. If this error is recurrent, generate a System
error report and contact service.
Unclean DropSync sensor
Clean the drop sensor photocell located above the
tube sensor with a damp cloth.
The tubing is not correctly mounted in the tubing
holder nut
Check that the tubing is correctly mounted, see Con-
nect tubing to the ÄKTA avant instrument, on page 134.
The fraction collector movement is blocked
Make sure that the fraction collector can move and is
free from obstructions.
The flow rate is too high
Reduce the flow rate.
The fraction volume is too small
Collect larger fractions.
Too many fraction collector instructions have been
sent
Wait for a while and try again.
8 Troubleshooting
8.5 Troubleshooting: Fraction collector
395