Step
5
6
Run a System CIP method
Follow the instructions to run a System CIP method.
Step
1
2
3
4
5
ÄKTA avant User Manual 29035184 AE
Action
In the Method Editor module,
click the Save the method button
or
on the File menu, click Save As
Result: The Save As dialog opens.
In the Save As dialog:
Select a target folder to enable the Save button.
Type a Name for the method.
Select a System from the list.
Click Save.
Result: The created method is saved in the selected folder.
Action
In the Method Editor module, create a System CIP method according to the
previous instructions.
Connect by-pass tubing to all selected column positions and loop positions
if a loop valve is used.
Prepare cleaning solutions and immerse the selected inlet tubing in the so-
lutions.
Note:
Note that each phase uses one solution. All inlets selected in one phase should
be immersed in the same cleaning solution.
In the System Control module, select the created method and start the run.
For complete cleaning of the flow path, clean the manual injection port of
the Injection valve and the pH valve manually, see the instructions below.
7 Maintenance
7.6 Maintenance when required
7.6.2 Perform System CIP
293