Edit system settings
It may be necessary to edit the System Settings when the configuration of the system
is changed. For example, if the change in configuration affects the delay volume following
the UV monitor (or other monitor connected via the I/O-box) the appropriate system
settings for Delay volumes have to be updated. This is to ensure that the fractions
marked in the chromatogram corresponds to the actual collected fractions.
Other system settings might also need to be edited for some optional modules.
All system settings available for ÄKTA avant are found in Section 9.7 System settings, on
page 459.
ÄKTA avant User Manual 29035184 AE
4 Optional instrument modules
4.2 Installation of optional modules