4 Optional instrument modules
4.12 I/O-box
4.12.4 Connect external equipment to the I/O-box
System properties
Follow the instruction to update the system properties.
Step
1
2
3
System settings
Default values for digital out ports, noise reduction and configuration of analog out ports
can be set.
Instruction name
Digital out X
Noise reduction analog
in X
Alarm analog in X
Alarm digital in X
Configure analog out X
Note:
152
Action
Open the system properties Edit dialog box.
In the Component types list, click Other.
Select the I/O-box (E9) or I/O-box 2 (E9) check box in the Component selec-
tion list. Then click OK.
Description
Sets the value of the signal sent out by digital port
number X to either 0 or 1. The default value is 1.
Filters the noise in the analog signal in port number X.
Enables or disables the alarm for the analog signal in
port number X. When enabled, it sets the alarm limits
for the analog signal. If the alarm is enabled and the
analog signal falls outside the set limits, an alarm will
be triggered and the method will be paused.
Enables or disables the alarm for the signal in digital
port number X. The alarm can be triggered by either of
the signal values, 0 or 1. If the alarm is enabled and the
condition set in 'Value' occurs, an alarm will be triggered
and the method will be paused.
Enables the user to send one of the pre-defined signals
(UV signal, conductivity, temperature, pH or concentra-
tion of eluent B) to the analog out port number X, and
also to set the range of that signal.
The delay volume has to be updated if an external component is added to the
flow path.
ÄKTA avant User Manual 29035184 AE