Clean Built-In Fraction Collector Sensors - GE AKTA avant User Manual

Hide thumbs Also See for AKTA avant:
Table of Contents

Advertisement

7 Maintenance
7.3 Weekly maintenance

7.3.3 Clean built-in fraction collector sensors

7.3.3
Clean built-in fraction collector sensors
Maintenance interval
Clean the fraction collector sensors every week, or when required, for example if the
fraction collector fails to read the tray ID or does not collect fractions correctly.
Required material
The following material is required:
Instructions
Follow the instructions to clean the fraction collector sensors. See Illustration of the dis-
penser head, on page 88 for the location of the fraction collector sensors.
Step
1
276
Wash bottle
Water or 20% ethanol
Cloth
Action
In the System Control module, on the Manual menu, click Execute Manual
Instructions. In the Manual Instructions dialog, select Fraction collec-
tion:Frac cleaning position and then click Execute.
Result: The dispenser head moves to cleaning position, and the instrument
display states System pause.
ÄKTA avant User Manual 29035184 AE