7.8.5

Replace UV monitor U9-L flow cell

Maintenance interval
Replace the UV flow cell when it is desired to use a flow cell with a different path length,
or if the cell is damaged.
Required material
Replace the flow cell
Follow the instruction to replace the UV flow cell.
Step
1
2
ÄKTA avant User Manual 29035184 AE
UV flow cell
CAUTION
Hazardous chemicals or biological agents in UV flow cell. Make
sure that the entire flow cell has been flushed thoroughly with
bacteriostatic solution (e.g., NaOH) and distilled water, before service
and maintenance.
Action
Switch off the instrument.
Disconnect the tubing from the UV flow cell.
7 Maintenance
7.8 Replacement procedures

7.8.5 Replace UV monitor U9-L flow cell

345