7.6.4

Clean built-in fraction collector

Maintenance interval
Clean the Fraction collector when required, for example if liquid has been spilled in the
Fraction collector chamber. The internal tubing of the fraction collector may need to be
replaced for maintenance or for process purposes. Information on how and when to
replace the internal tubing can be found in Section 7.8.1 Replace tubing and connectors,
on page 336
Required material
The following materials are required:
Instruction
Follow the instruction below to clean the interior of the fraction collector. See the locations
of the components of the fraction collector in Section 3.10.2 Illustrations of the built-in
fraction collector, on page 86.
Step
1
ÄKTA avant User Manual 29035184 AE
Wash bottle
Water or 20% ethanol
Cloth
Action
In System Control module, on the Manual menu, click Execute Manual In-
structions.
Result: The Manual instructions dialog opens.
7 Maintenance
7.6 Maintenance when required

7.6.4 Clean built-in fraction collector