7 Maintenance
7.6 Maintenance when required
7.6.3 Perform Column CIP
Step
7
8
Run a Column CIP method
Follow the instructions below to run a Column CIP method.
Step
1
2
3
4
300
Action
In the Method Editor module,
click the Save the method button
or
on the File menu, click Save As
Result: The Save As dialog opens.
In the Save As dialog:
Select a target folder to enable the Save button.
Type a Name for the method.
Select a System from the list.
Click Save.
Result: The created method is saved in the selected folder.
Action
In the Method Editor module, create a Column CIP method according to
the previous instruction.
Prepare cleaning solutions and immerse the selected inlets in the solutions.
Connect the column to the selected column position.
In the System Control module, select the created method and start the run.
ÄKTA avant User Manual 29035184 AE