6.3 Chemcassette™ Analyzer Maintenance
This section describes maintenance and handling procedures for theChemcassette™ analyzer.
Unless noted otherwise, these procedures apply to both the Universal Chemcassette™ and the
pyrolyzer model analyzers.

6.3.1 Remove and Replace Analyzer Particulate Filters

The Vertex M Chemcassette™ analyzer uses various filters to protect the unit from particles and
potentially damaging gases. Table 5-1 provides maintenance information about filters.
The Vertex M Analyzer houses filters in a filter magazine. Use the following procedure when
replacing filters.
1. Set analyzer to idle in the Runtime Options window.
2. Touch Main Screen, Menu, Maintenance and then Change Filter.
3. Touch the Analyzer module number in the Analyzer selection keypad.
4. Touch Release Analyzer.
5. Slide analyzer out of cabinet until filter compartment is visible.
Vertex M
121
User Manual