6.2 Maintenance Schedules
Perform maintenance following the schedule in Table 6-1. See "Replacement and Consumable
Items" on page 177 for part numbers of maintenance items.
Teflon Corrosive Filter Membrane (end of Iine)
Cabinet filter (located in front of pump module)
Proportional valve filter (inside analyzer)
Vertex M
Item
Sample line filters (end of line)
Teflon Corrosive Filter
Pump vane replacement
Pump stem and o-ring
Particulate Filters
Pyrolyzer Freon Filter
Acid Scrubber Filter
Alternate Pumps
Optics Cleaning
System File Maintenance
Table 6-1. Suggested Maintenance Schedule
3 months or as needed
2 years operation per pump
120
Schedule
3-6 months
1 month
3 months
6 months
1 year
3-6 months
1 month or as needed
6 months
6 months
1 year or as needed
1 year or as needed
User Manual