Honeywell Vertex M User Manual page 195

Continuous gas monitor
Hide thumbs Also See for Vertex M:
Table of Contents

Advertisement

Tag
Folder
Count
DtoA
72
Flow
72
Pressure
72
Slope
72
9
System
9
72
72
GasCon
72
Main
0
Ana_Fault
9
PLC
9
Fault
72
Vertex M
Recommended
for most HMI
systems
##
##
##
##
#
Az#
Pt##Flow10
Pt##Flow90
Y
##1
Y
#
Y
#
Y
##
195
Tag Name
Value Interpretation / Descrip-
Format
For HA internal use only
Integer sample point flow in
cc/minute
Float32, point sample pressure
in inches Hg, (usually approx. -
1.0)
Float32 for HA internal use only
Float32, vacuum pump pressure
in inches Hg, as measured by
each analyzer (usually -13.0)
String in form "Az x-y"
Float32 for HA internal use only
Float32 for HA internal use only
Float32; Gas concentration in
ppm, ppb, mg/m
Unsigned integer Status
0 = No fault
1 = Maintenance Fault
2 = Instrument Fault
3 = Both
Reports only analyzer-wide
faults. Does not indicate the
presence of a point-specific
fault
Unsigned integer Status
0 = No fault
1 = Maintenance Fault
2 = Instrument Fault
3 = Both
Reports faults from either
analyzer or point
Unsigned integer Status
0 = No fault
1 = Maintenance Fault
2 = Instrument Fault
tion
3
as configured.
User Manual