Honeywell Vertex M User Manual
Honeywell Vertex M User Manual

Honeywell Vertex M User Manual

Continuous gas monitor
Hide thumbs Also See for Vertex M:
Table of Contents

Advertisement

Quick Links

Vertex™ M
Continuous Gas Monitor

Advertisement

Table of Contents
loading

Summary of Contents for Honeywell Vertex M

  • Page 1 Vertex™ M Continuous Gas Monitor...
  • Page 2: Table Of Contents

    Chapter 2 - Introduction 2.1 Trademarks 2.2 System Overview 2.2.1 Manufacturer 2.2.2 General Safety 2.2.3 Vertex M Series Rack and Analyzer Definitions 2.3 System Components 2.3.1 Vertex M Front 2.3.2 Vertex M Back 2.3.3 Sample Tubing, Exhaust, and Wiring Ports 2.3.4 Analyzer Side Panel...
  • Page 3 2.11.1 Data Acquisition Computer 2.11.2 Programmable Logic Controller Chapter 3 - Installation 3.1 Introduction 3.2 Surveying the Installation Site 3.2.1 Placement of the Vertex M system 3.2.2 Exposure to Dust and Humidity 3.2.3 Sample Transport Time 3.2.4 Monitor Dimensions 3.2.5 Sample Locations 3.2.6 Sample Line Particulate Filter Use...
  • Page 4 4.6.10 Profile Management-File Menu 4.6.11 Other Menu 4.7 Load Tape 4.8 Verify Flow Rates and Supply Vacuum 4.8.1 Verify Flow Rates 4.9 Leak Checking Sample Lines 4.10 Reconfigure 4.11 Moving to a New Site 4.12 System Shutdown Vertex M User Manual...
  • Page 5 4.12.1 Vertex M system with Windows 10 4.12.2 Vertex M system with Windows 7 Chapter 5 - Operation 5.1 Introduction 5.2 Monitoring Mode Overview 5.3 Main Screen 5.3.1 System Display Area 5.3.2 Point Detail Display Area 5.3.3 Function Buttons 5.4 Project Functions 5.4.1 Log In and Log Out...
  • Page 6 6.7.1 Remove Supply 6.7.2 Replace Supply 6.8 Clean the Touch Screen 6.9 PLC Module Battery Backup Check 6.10 File Maintenance 6.10.1 Vertex M with Windows 7 6.10.2 Vertex M with Windows 10 6.11 Optics Cleaning Chapter 7 - Troubleshooting 7.1 Introduction 7.2 General System Problems...
  • Page 7 8.6.1 Relay Output Contacts 8.6.2 Relay Contact Ratings 8.7 Series 1 Rack Network Interface and Options 8.7.1 OLE for Process Control (OPC) Interface 8.7.2 Data Values Common to Fieldbus Networks 8.7.3 Data Map 8.7.4 Profibus Option (P/N 1295-275) Vertex M User Manual...
  • Page 8 8.7.8 Modbus Plus Interface (P/N 1295-0330) 8.7.9 LonWorks Interface (P/N 1295-0329) 8.7.10 Modbus/TCP (P/N 1295-0520) 8.7.11 Ethernet/CIP (P/N 1295-0519) 8.7.12 Vertex M Remote Display Setup 8.7.13 Demonstrating Remote OPC Communication 8.8 Series 3 PLC Network Interface and Options 8.8.1 Common Considerations of Anybus Communicator 8.8.2 Common Considerations of Anybus X-Gateways...
  • Page 9 8.11.3 Software Overview 8.11.4 Accessing LIT Parameters 8.11.5 LIT Access from the HMI 8.11.6 LIT Related Events 8.12 Warranty Statement 8.12.1 Chemcassette™ Device Warranty Statement 8.12.2 Chemcassette™ Warranty 8.13 Contact Us Vertex M User Manual...
  • Page 10: Chapter 1 - Safety

    WARNING The most dangerous or potentially lethal hazards: unguarded equipment, confined space entrances, and lockout labels. DANGER Caution: possibility of electric shock Caution: hot surface Protective conductor terminal (ground terminal) Vertex M User Manual...
  • Page 11: Emc Considerations

    1.2 EMC Considerations Honeywell's monitor has been designed to comply with applicable Electromagnetic Compatibility (EMC) standards at the time of manufacture. The design includes filtering, shielding, and bypassing techniques. At the time of certification, simulated customer Input/Output (I/O) schemes were tested. All methods used in the equipment for emission suppression and reduction of susceptibility are interactive.
  • Page 12: Cabling

    (absolutely no greater than three inches or eight centimeters). For multiconductor connector terminations, only 360° shielded shells should be used. Note: Honeywell product testing uses >90% braid with foil (around the bundle); twisted pair; stranded 24 AWG (minimum wiring for all qualification and certification testing.) 1.6 Connectors All qualification and certification of Honeywell products were achieved with high quality connectors, providing 360°...
  • Page 13: Chapter 2 - Introduction

    Storing the alarm information in a database. The Vertex M system provides fast response to a wide range of gases. Each location may be up to 400 ft (122 m) from the Vertex M system. The system uses one or more of Honeywell’s Chemcassette™...
  • Page 14: Manufacturer

    Follow all installation and operational instructions to ensure the safe and reliable operation of this unit. If this monitor is used in a manner not specified by Honeywell, the protection provided by the equipment may be impaired. 2.2.3 Vertex M Series Rack and Analyzer Definitions Series 1 –...
  • Page 15: System Components

    2.3 System Components The following photos illustrate Vertex M system components, ports, connections and controls. From the main front and back photos, click on the labels to see the detail photos. 2.3.1 Vertex M Front Item Item Sample Tubing, Exhaust, Universal Chemcassette™...
  • Page 16: Vertex M Back

    2.3.2 Vertex M Back Item Item Data Acquisition Computer Pyrolyzer step-up/Isolation (rear) transformer 4-20mA Option PLC (Series 1 Back of Pyrolyzing unit only) Chemcassette Analyzer Vertex M User Manual...
  • Page 17: Sample Tubing, Exhaust, And Wiring Ports

    2.3.3 Sample Tubing, Exhaust, and Wiring Ports Note: The Alarm wiring conduit plate must remain in place if not used. 2.3.4 Analyzer Side Panel Exterior View Vertex M User Manual...
  • Page 18: Analyzer Front

    Internal View 2.3.5 Analyzer Front Vertex M User Manual...
  • Page 19: System Controls

    2.3.6 System Controls 2.3.7 Data Acquisition Computer (rear) Series 1 Rack Vertex M User Manual...
  • Page 20 Series 3 Rack CAUTION Restrict access to the USB port to reduce the risk of malicious software being introduced. Note: This photograph shows a typical port configuration. Port and slot locations vary from model to model. Vertex M User Manual...
  • Page 21: Back Of Chemcassette™ Module

    2.3.8 Back of Chemcassette™ Module *Note: Connection secured by slide latch. Push up to open. Push down to close. Vertex M User Manual...
  • Page 22: Main Plc

    2.3.9 Main PLC Series 1 Rack Series 3 Rack Vertex M User Manual...
  • Page 23: 4-20Ma Analog Output Option Plc

    2.3.10 4-20mA Analog Output Option PLC Series 1 Rack Series 3 Rack Vertex M User Manual...
  • Page 24: Menu Map

    2.4 Menu Map Vertex M User Manual...
  • Page 25: Analyzer Modules

    Sample atmosphere runs from the location to the Vertex M system via a line Each of the 24 sample tubing connections on the Vertex M system corresponds to a point. A sample line can be connected directly to a single point or multiple points via a 4-port manifold (PN: 1295-0349) The system draws air simultaneously from all locations.
  • Page 26: Chemcassette™ Detection System

    2.7 Chemcassette™ Detection System The Chemcassette™ Analyzer module is a self- contained, microprocessor controlled analyzer that occupies one slot in a Vertex M. Sample lines and the vacuum source are connected to the Chemcassette™ via a single 10-tube connector. The system powers up in the same state as when powered down. Data is stored in the module’s memory until the data acquisition computer retrieves it.
  • Page 27: Detector Optics

    Chemcassette tape to the sample outlet. The target gas in the sample flow reacts with the tape and produces a stain density proportional to the gas concentration. An LED in the detector head illuminates the sample stain and the detector then optically measures the stain. Vertex M User Manual...
  • Page 28 LED. The microprocessor in the Chemcassette analyzer module interprets the stain. It then calculates and stores a precise concentration level in the module’s memory. Gas concentrations are reported in parts-per-million (ppm), parts-per-billion (ppb) or milligrams-per-cubic-meter (mg/m Vertex M User Manual...
  • Page 29: Chemcassette™ Tapes

    Particulate filters protect the internal precision orifice from dust particles. An acid filter is used on the common line to the pumps. Both types of filters are located in a removable filter block on the side of the Chemcassette™ module. An internal particulate filter protects each proportional valve. Vertex M User Manual...
  • Page 30: Pyrolyzer Module Detection System

    Charcoal filters have a part number (P/N 1874-0139) unique to the pyrolyzer module. The Vertex M Pyrolyzer requires two adjacent slots on one tier and always occupies Slot 1 and 2. The bottom rail and latch must be removed from slot 1 to install pyrolyzer.
  • Page 31: Vacuum Pumps

    The pumps are located in the bottom of the Vertex M system cabinet inside a sound-deadening enclosure to reduce noise. Three cooling fans circulate air over the pumps. The Vertex M system draws cooling air in through a filter mounted on the pump module access door.
  • Page 32: Control System

    2.11 Control System The Vertex M control system is a redundant system consisting of a central data acquisition computer (DAq), a programmable logic controller (PLC) and one or more analyzer modules. Figure 2-3. Communications Path Above is a simplified block diagram of the communications path of the control system. The analyzer modules and PLC are microprocessor controlled and contain non- volatile memory 2.11.1 Data Acquisition Computer...
  • Page 33: Chapter 3 - Installation

    CHAPTER INSTALLATION 3.1 Introduction The installation and initial start-up procedure for the Vertex M system consists of seven steps, described in this and the following sections: Surveying the Installation Site Optional Floor Mounting Installing Sample Lines/Filters Installing Pump Exhaust Line...
  • Page 34: Placement Of The Vertex M System

    Note: Refer to "Installation Drawings" on page  157 on lifting/mounting. The Vertex M system can be placed at maximum 400 ft. See "Sample Transport Time" below for more information. 3.2.2 Exposure to Dust and Humidity Exposure to corrosive gases or materials, excess moisture, dust and other unusual environmental conditions could seriously hamper the Vertex M’s monitoring ability and could...
  • Page 35: Sample Line Particulate Filter Use

    ). Before enabling building alarm systems, make certain that 1) the correct Chemcassette™ is installed and 2) the monitor reads zero. Install sample lines from each location to the top of the Vertex M system. This procedure involves: Sample Line Installation Requirements...
  • Page 36: Sample Line Connections

    Unused sample line ports should have a particulate filter installed to keep the system clean. If an analyzer is installed in the Vertex M with a Chemcassette tape, the optics may need cleaning before activating a previously unused point.
  • Page 37: Installing Sample Line Particulate Filters

    This section describes exhaust connections and installation. The Vertex M system is equipped with a vacuum pump that is located in the bottom of the Vertex M system cabinet. The pump exhaust line connects to the manufacturing facility central toxic exhaust system.
  • Page 38: Exhaust Line Connection

    3. Cut the tube end perpendicular to its length 0.062 inches (1.5 mm) from its end. 4. Insert the tube in the fitting to a depth of 0.95 in. (24.13mm) ±0.05 inches (1.27mm) With the system running, verify the leak integrity with a small amount of leak test fluid. Vertex M User Manual...
  • Page 39: Electrical Power

    Operating Voltage: 110 or 230 VAC ± 10% (under load) @ 50/60 Hz; 15 Amps maximum, single phase. The Vertex M system requires a dedicated AC circuit rated at either 110 or 230 volts, 50/60 Hz, 15 Amp single phase. Line voltage should fluctuate no more than ± 10%. The external switch must be clearly labeled and installed in accordance with local electrical codes.
  • Page 40: Power On/Off

    3.6.2 Power On/Off An internal rack power switch is located behind the door. After performing self-diagnostics, the Vertex M system main screen opens and the system returns to the same state it was in prior to power down. Figure 3-3. Rack Power Switch...
  • Page 41: Data Acquisition System

    3.7 Data Acquisition System The data acquisition computer or DAq is the main computer in the Vertex M system. System display and operator control are through an LCD touch screen with on-screen keyboard or the external keyboard on top of the unit.
  • Page 42: Printer

    3.7.1 Printer The Vertex M system software can be programmed to print to either a network or local printer. To install a local printer, connect it to the parallel printer port as shown. The USB port may also be used. The correct printer driver must also be installed.
  • Page 43: External Network Connection

    3.7.2 External Network Connection The Vertex M system can be connected to an external Ethernet network at the port shown. Series 1 Unit Figure 3-6. Series 3 Unit Figure 3-7. Vertex M User Manual...
  • Page 44: Network Computer Security

    Figure 3-8. Ethernet cable When using an Ethernet output, the conduit must be connected to Earth ground. The Vertex M comes with a Ferrite. Four wraps of the Ethernet cable is needed as close to the Vertex M as possible.
  • Page 45: Wiring Alarm Relays

    3.8.1 Relay Contacts The Vertex M system has form-A, single-pole, single- throw relays that activate external alarm devices. Contacts are available for each circuit to accommodate installation of external devices. Relay panels are located behind theVertex M LCD screen. See "Series 1 Rack Optional Relay Specifications"...
  • Page 46 Do not switch DC current with the relay contact unless counter electromotive force (CEMF) protection is being used such as a suppression diode. Do not use the Vertex M system power supply for external alarm power. Note: Make sure all connections comply with applicable RFI/EMI standards.
  • Page 47: Chapter 4 - Startup

    This section describes the Vertex M system startup sequence. 4.1 Initial Startup Use this section to turn on the Vertex M system and to configure the analyzer modules for specific gas locations. There are six parts to this startup procedure:...
  • Page 48: Verify Installation

    Relay wiring See "Installation" on page  33 for connection details. 4.4 Power Up Use the rack power switch behind the front door to power up the Vertex M system. 1. Open front door. 2. Turn on rack power switch. 3. Turn on power switch to appropriate analyzers.
  • Page 49 Table 4-1. Analyzer Status LEDs Vertex M User Manual...
  • Page 50: Start Program

    Note: Use the Windows Date/Time Properties dialog box to change the time zone, time and date on the Vertex M system. Stop project if adjusting time and time zone. Once complete, restart project. See "Stopping Project" on page  91 for more information on how to stop the project.
  • Page 51: Configuration Utility

    4.6 Configuration Utility Before the Vertex M system can begin monitoring, a configuration profile must be created. The configuration profile stores all of the monitor settings in a single file on the hard drive. Configuration profiles include system level information, point settings and analyzer information.
  • Page 52 Vertex M User Manual...
  • Page 53 Vertex M User Manual...
  • Page 54 Vertex M User Manual...
  • Page 55 Note: Setting the Vertex M system to continuously log concentration data on a 3-analyzer system requires approximately 15 megabytes of disk storage per day at the fastest logging rate of once every 5 seconds. Purge data often to avoid filling available disk space.
  • Page 56: Database Management-Retention Periods

    1-99. Vertex M will not recalculate the values when the unit is changed. For example, if the purge period is 14 days and “days” to “weeks” is changed, Vertex M will set the period to 14 weeks. Vertex M performs the record purge as the data acquisition computer clock passes midnight.
  • Page 57: Define Gas Location

    4.6.2 Define Gas Location Use Define Gas Location to edit the list of locations. Assign a long and a short name for each location. Vertex M User Manual...
  • Page 58 Vertex M User Manual...
  • Page 59: Configure Analyzers And Points

    4.6.3 Configure Analyzers and Points Press “Configure Analyzer/Points” to change the right side of the Configuration window to a display representing physical layout of the Vertex M system. Each slot is represented by a two- part button. When the analyzer has been configured, the top of the button displays the type of analyzer and the gas family.
  • Page 60: Configure Point

    ChemCam AutoPicture When either ChemCam for Level 1 or Level 2 gas is selected, a gas alarm signals the Vertex M system to store a picture of the stain the next time it advances the Chemcassette™ tape. When the tape advances, the tape stops with the stain under the camera, the ChemCam takes a picture and then advance continues.
  • Page 61 169 for a list of default LDL levels. The user-configured LDL can be disabled by un- checking the check box. When the user-configured LDL is disabled, the Vertex M system will use the default LDL level. A new level may be entered only within the range of the Vertex M detection system.
  • Page 62 Note: Setting the Vertex M system to continuously log concentration data on a 3-analyzer system requires approximately 15 megabytes of disk storage per day. Purge data often to avoid filling available disk space. Vertex M...
  • Page 63 Units Selects the unit of measure to display target gas concentrations. This selection applies to this screen only and does not effect displays during normal operation or events. Vertex M User Manual...
  • Page 64 Vertex M User Manual...
  • Page 65 XP Hydrides 15-25 Mineral Acids 30-40 /Oxidizers (NO 30-40 Mineral Acids (Pyro) 30-40 Phosgene 30-40 XP Phosgene 30-40 Aliphatic Amines 15-25 /Oxidizers-III 30-40 Hydrogen Cyanide 15-25 Hydrogen Sulfide 15-25 XPV Chlorine 30-40 XPV Chlorine-II 15-25 Fluorine/Oxidizers 15-25 Vertex M User Manual...
  • Page 66 XPV Mineral Acids (Pyro) 30-40 LL Sulfur Dioxide 15-25 XP4 Hydrides 15-25 XP4 Mineral Acids 30-40 XP4 Mineral Acids (Pyro) 30-40 XP4 Chlorine 30-40 XP4 Ammonia 15-25 XP4 Phosgene 30-40 XPV Germane 15-25 XPV Chlorine-3 15-25 Germane 15-25 Vertex M User Manual...
  • Page 67: Define And Assign Relays

    4, 8, or 16 relays populate each card. Note: Relays are NOT factory configured. Vertex M uses a two-part display to configure relays. The Configure PLC window displays the available relay cards and contacts. In a second window, Vertex M also displays a representation of the analyzers installed in the system.
  • Page 68: Configure Plc

    PLC Relay Cards Up to 4 relay cards populate the Series 1 Vertex M System. The cards are numbered 3-6. In case of Series 3 Vertex M System, up to 8 cards can be populated and the cards are numbered 2-9.
  • Page 69 When the number of fault relays is changed, Vertex M opens the Fault Relay Definition Change dialog box. Choosing Yes shifts the alarm definitions, choosing No pads the alarm definitions.
  • Page 70: Set Alarm Relays

    To verify a contacts association to analyzer points, choose the contact. The associated point display changes color. Note: Number of card slots vary based on PLC configuration such as number of relay cards and relays per card. Vertex M User Manual...
  • Page 71: Set Fault Relays

    Instrument faults indicate a loss of monitoring on one or more points. Maintenance faults indicate the Vertex M system requires attention but is continuing to monitor. Note: The number of relays used for fault indication is configurable. See "Configure PLC" on page ...
  • Page 72: Profile Management-File Menu

    Use the file menu to open, create, save or close a configuration file. Configuration files may be stored in any directory on the Vertex M system hard disk. When any changes are made to the configuration profile, Vertex M will always prompt the user to save the change before closing the Configuration window.
  • Page 73: Load Tape

    The Authorized Service window displays the flow range in bar graph form for each point of the selected analyzer. Choose the analyzer from the selection pad in the upper right-hand corner. Press the pump on button. The eight points display their flow. Vertex M User Manual...
  • Page 74 Vertex M User Manual...
  • Page 75: Leak Checking Sample Lines

    To troubleshoot the condition, disconnect the sample line at the inlet port at the top of the Vertex M cabinet. Securely plug the inlet port and repeat the above leak check procedure. If the sample point passes the test with the top port plugged, the leak is somewhere in the sample line and the line must be replaced.
  • Page 76: Reconfigure

    4.11 Moving to a New Site Before moving the Vertex M to a new site, use the following procedures to prevent loss of data or damage to the monitor. 1. Remove all Chemcassette™ tapes and store as required by local policies.
  • Page 77: System Shutdown

    Crate and pad the Vertex M to prevent damage during transport. If unsure of packing requirements, contact the Honeywell Service department. CAUTION Leaks in the exhaust tubing connection can cause exposure to toxic gases from remote sample areas. For leak-tight connections, follow the instructions in the Installing Pump Exhaust Line section and the Remove Pump section.
  • Page 78: Vertex M System With Windows 7

    4.12.2 Vertex M system with Windows 7 1. Exit the Vertex M program. Touch Project and then Stop Project. 2. When the Vertex HMI has closed, stop RSView32. Vertex M User Manual...
  • Page 79 Note: RSLinx must be shut down before closing the OPC server. Failure to do this will result in the OPC server being automatically restarted. 5. In the Windows taskbar touch Start and then Shut Down. 6. Open touch screen and set all switches and the rack power switch to Off. Vertex M User Manual...
  • Page 80: Chapter 5 - Operation

    Menu Buttons OnScreen Keyboard See "Startup" on page  47 if the analyzers in the Vertex M system have not yet been configured. 5.2 Monitoring Mode Overview Monitor mode is the Vertex M system’s standard operating state. Upon power up, the monitor performs initialization routines and returns to the same state as when powered down.
  • Page 81: Main Screen

    Point detail screen Optional fieldbuses Optional 4-20 mA output Data logger 5.3 Main Screen The Vertex M system opens the main screen after power up. Vertex M divides the main screen into three areas: System display Point detail Function buttons...
  • Page 82: System Display Area

    5.3.1 System Display Area The System Display Area displays information about all of the points in the Vertex M system. Each of the 72 blocks in the System Display Area represents one point. A group of eight blocks represents an analyzer block. The analyzer block has the status bar indicator at the top of the block.
  • Page 83 FLOW - a flow Auto Balance procedure is in process on the analyzer The Vertex M system displays only the short names of the location and target gases within each point block. During normal monitoring, the background color of each block is white. Vertex M will change the background color of a point as conditions change.
  • Page 84 Pump Status Indicator The Vertex M Display includes pump status indicators in the top right corner. The bottom row indicates which of the pumps is currently operating. The indicator will display ON with a green background if the pump is operating and OFF on a white background if the pump is not running.
  • Page 85: Point Detail Display Area

    Points not needed for monitoring may be disabled. A disabled point does not Point trigger alarms See "Event History" on page  92 for more information. See "Data Trend" on page  97 for more information. See "Event List" on page  100 for more information. Vertex M User Manual...
  • Page 86 Figure 5-1. Figure 5-2 is an image of the Point Detail section of the main Vertex M display screen snapshot. Analyzer The analyzer type Points / Alarms A point may be enabled but the alarm may be disabled Name of the gas or gases, up to three gases per location...
  • Page 87: Function Buttons

    Review - View stored historical gas concentration data or events and access ChemCam. Project - Log in, log out, change passwords, update programs and stop the project (exit Vertex Help - Opens a window to explain the functions of the main screen. Vertex M User Manual...
  • Page 88: Project Functions

    5.4.1 Log In and Log Out To protect the integrity of the system, the Vertex M system classifies menus as either open or protected functions. If access to a protected menu is required, log in under a user account with permission to use that menu.
  • Page 89 The Vertex M system will automatically log out any user after a period of inactivity. The default timeout period is 8 hours. Authorized users may change the timeout setting in the Configuration Menu. Thirty seconds prior to the end of the timeout period, Vertex M will warn before logging out a user. Vertex M...
  • Page 90: Changing Password

    4. Enter new password in both text boxes. 5. Touch Enter. If both new passwords are identical, Vertex M will accept the new password. Change Password – RSView32 HMI / Windows 7 Change Password – FactoryTalk View HMI / Windows 10 5.4.3 Updating Program...
  • Page 91: Restore Onscreen Keyboard

    If keyboard is hidden, restore it with the OnScreen Keyboard button from the Project Menu. 5.4.5 Stopping Project Use Stop Project to exit the Vertex M program. Touch Project and then Stop Project. Even though the Vertex M program is not running, individual analyzers continue to monitor and store data in their internal memory.
  • Page 92: Event History

    5.5.1 Event History An event is any action that the Vertex M system is required to enter into the database. As events occur, the Vertex M system stores the events in a database on the DAq PC. The default database format is Microsoft Access.
  • Page 93 Figure 5-4. Event History Figure 5-5. Event History Vertex M User Manual...
  • Page 94 Figure 5-6. Event History Vertex M User Manual...
  • Page 95 Advanced Filter Options-Events Use to filter on specific alarms, faults, or other events. Figure 5-7. Filter Options Vertex M User Manual...
  • Page 96 Use to filter events based on specific points, analyzers or other hardware modules. Figure 5-8. Filter Options Select User ID or Logging Station Use to select a specific Vertex M station or operator. Figure 5-9. Selecting Station or ID Vertex M...
  • Page 97: Data Trend

    5.5.2 Data Trend Vertex M plots gas concentration in the Data Trend window. Multiple points may be chosen to plot each point on a different colored line on the graph. Data displayed in the graph can be either Stored data from the data log or Live (a continuous, real-time display, scrolling right to left).
  • Page 98: Optional Chemcam

    5.5.3 Optional ChemCam The Vertex M ChemCam option provides a means to observe the stains on the last window of a Chemcassette™ tape. It consists of a small video camera located between the Chemcassette™ take up reel and the optic head. The Vertex M system records images of the stains in either AutoPicture mode or by a ChemCam Live capture.
  • Page 99 Additional information for records stored by ChemCam AutoPicture includes Point number which triggered the alarm Gas concentration at the time of alarm Unit of measure for the concentration The alarm level the concentration reached The location ID Figure 5-12. ChemCam Database Vertex M User Manual...
  • Page 100: Event List

    5.5.4 Event List With the Event List, view and acknowledge active alarms and faults. Vertex M displays the newest event at the top of the list. Scroll through the events with the Back and More buttons. To select an event as the current event, touch the event. The current event will have a green border.
  • Page 101: Run Time Options

    The event will occur after the timeout period set in the configuration window. Vertex M always displays the following items in the Run Time Options window. To Start or Stop an Analyzer from Monitoring Stopping an analyzer from monitoring prevents the system from triggering false alarms during testing and maintenance.
  • Page 102 To Disable or Enable a Point Disabling a point prevents the system from triggering false alarms during testing and maintenance. Vertex M will not perform point specific fault checks on disabled points. 1. Choose Point Enable/Disable from the function buttons.
  • Page 103 Choose All Points for Location to enable / disable all of the alarms associated with a location. 3. Touch Alarm Enable or Alarm Disable to change the state of the point. A disabled alarm turns light blue. An enabled alarm turns green in the point status window. Vertex M User Manual...
  • Page 104: Flow Calibration

    Comment Event Use Comment Event to enter information about a point or analyzer that Vertex M would not automatically enter into the database. The comment is stored in Event History. To enter a comment: 1. Touch Comment Event. 2. Enter the text for the event.
  • Page 105 Function buttons Flow Display The flow display shows information critical to verifying the flow rate for each point. Choose the analyzer from the selection pad in the upper right-hand corner. The eight points display their flow. Vertex M User Manual...
  • Page 106 Function Buttons Vertex M User Manual...
  • Page 107 Autobalancing takes approximately 130 seconds to complete. 6. If required, repeat for other analyzers. Note: If Vertex M cannot adjust the flow to the proper range (150 to 210 cc per minute at worst case condition) consult Honeywell service. Flow Calibration During Auto Balance...
  • Page 108: Maintenance

    1. Touch Main Screen, Menu, Maintenance then Analyzer Operations. 2. Touch the module number on the module keypad. 3. Select Gate Open. 4. When finished, press Gate Close and return analyzer to monitor mode in Runtime Options window. Vertex M User Manual...
  • Page 109: Diagnostics

    The Profile ID is a unique number generated whenever the Vertex M Profile Management utility saves a configuration file. Vertex M stores the ID number in the DAq, the PLC and each analyzer. If the profile ID numbers do not match, a fault is generated and monitoring will not start.
  • Page 110 PLC Information Displays the profile ID number in the PLC. In case of Vertex M with Series 3 Rack PLC system, the program version will be displayed. Vertex M User Manual...
  • Page 111: Service

    Fault Test Use the Fault Test to verify operation of fault relays. Note: The fault test simulates an actual fault condition and the Vertex M system activates fault relays. Notify appropriate personnel that a fault test is planned to be conducted.
  • Page 112 Use the alarm test to simulate a gas concentration for any analyzer. Note: The alarm test simulates an actual alarm condition and the Vertex M system activates all alarm relays. Notify appropriate personnel that an alarm test is planned to be conducted.
  • Page 113 3. Use the up and down arrows to set the current level. Touch 20 mA Test to start the test. Touch Reset to stop the test. 4. To continue testing, repeat steps 2 and 3. See "Event List" on page  100 to clear events. Vertex M User Manual...
  • Page 114: Security Access

    5.6.6 Security Access RSView32 under Windows 7 Access to Vertex M functions is by a permission list which is stored in a user account. Only users authorized access to Security Setup may create or change user accounts. The system administrator assigns permission to Security Window and other protected functions by using the Security Setup menu.
  • Page 115 Unfortunately, Chemcassette™ installation inherently requires interrupting monitoring and requires human intervention. The Vertex M can be set up to issue a fault if this intervention is not performed promptly. See "Configuration Utility" on page  51 for more information. However, for thorough protection against errors by personnel who are only authorized to install Chemcassette™, many of the permission bits must be set to No.
  • Page 116: Configuration

    Windows operating system, where Engineers role is limited to the HMI only. The management of the roles is done outside of the Vertex M HMI and is part of the Rockwell software suite. Open the FactoryTalk Administration Console to manage roles. Please refer to the Rockwell’s FactoryTalk Security System Configuration Guide for usage details.
  • Page 117: Onscreen Keyboard

    5.7 OnScreen Keyboard The Vertex M display provides an onscreen keyboard for data entry. On-Screen Keyboard – RSView 32 / Windows 7 Vertex M User Manual...
  • Page 118: Restore Onscreen Keyboard

    On-Screen Keyboard – FactoryTalk View / Windows 10 5.7.1 Restore OnScreen Keyboard If keyboard is hidden, restore it with the OnScreen Keyboard button from the Project Menu. Vertex M User Manual...
  • Page 119: Chapter 6 - Maintenance

    Clean the Touch Screen PLC Module Battery Backup Check File Maintenance Optics Cleaning Most of the procedures in this section use functions in the Maintenance Window. To reach the Maintenance Window, touch Main Screen, Menu and then Maintenance. Vertex M User Manual...
  • Page 120: Maintenance Schedules

    Pyrolyzer Freon Filter 1 month or as needed Acid Scrubber Filter 6 months Alternate Pumps 6 months Optics Cleaning 1 year or as needed System File Maintenance 1 year or as needed Table 6-1. Suggested Maintenance Schedule Vertex M User Manual...
  • Page 121: Chemcassette™ Analyzer Maintenance

    6.3.1 Remove and Replace Analyzer Particulate Filters The Vertex M Chemcassette™ analyzer uses various filters to protect the unit from particles and potentially damaging gases. Table 5-1 provides maintenance information about filters. The Vertex M Analyzer houses filters in a filter magazine. Use the following procedure when replacing filters.
  • Page 122: Remove Analyzer Filters

    6.3.2 Remove Analyzer Filters Figure 6-1. Analyzer Filter Door Figure 6-2. Vertex M Filter Magazine 1. Open filter compartment by rotating the latch downwards and pulling handle down until door is at a 90˚ angle to the analyzer. 2. Remove filter magazine by grasping the magazine with thumb and middle finger and pulling straight out.
  • Page 123: Replace Analyzer Filters

    2. Position new filters in magazine with the filter flow direction arrow pointing down. Press into place. 6.3.3 Replace Analyzer Filters Figure 6-3. Analyzer Filter Door Figure 6-4. Vertex M Filters 1. Insert magazine into filter compartment. 2. Note orientation of arrow on side of magazine. 3. Close compartment door.
  • Page 124: Change Chemcassette™ Tape

    5. Thread tape leader. 6. Position leading alignment mark on Chemcassette™ tape under front edge of optic head. 7. Touch the NEXT button to verify optics. The tape advances as the Vertex M verifies the optics. 8. Push the analyzer into the cabinet.
  • Page 125: Replacing An Analyzer

    6.4 Replacing an Analyzer The Vertex M rack is designed for quick replacement of major components. Both the Chemcassette™ and pyrolyzer analyzers may be replaced while other analyzers continue to monitor. 6.4.1 Disconnecting Cables In steps 4 and 5 of the sequence that follows, the circular tubing harness and four electrical cables from the back of the analyzer need to be disconnected.
  • Page 126: Install Analyzers

    Return to Service 1. Open the front door and turn analyzer power switch (and pyrolyzer power switch if necessary) on. 2. Re-install Configuration Profile. 3. Install Chemcassette™. 4. Return analyzer to monitor mode in Runtime Options Menu. Vertex M User Manual...
  • Page 127: Remove And Replace Pyrolyzer Filters

    Heating nitrogen trifluoride converts it to hydrogen fluoride which the Vertex M pyrolyzer detects with a standard mineral acid Chemcassette™. The pyrolyzer heaters and associated control circuits are maintenance items not found in the Universal Chemcassette™...
  • Page 128: Remove Filters

    6.5.1 Remove Filters Figure 6-5. Pyrolyzer Filter Door Figure 6-6. Vertex M Pyrolyzer Filter 1. Open filter compartment by pulling handle down until door is at a 90˚ angle to the analyzer. 2. Remove filter magazine by grasping the magazine with thumb and middle finger and pulling straight out.
  • Page 129: Replace Filters

    1. Note orientation of arrow on side of magazine. 2. Insert magazine into filter compartment. 3. Close compartment door. Return to Service 1. Push analyzer into cabinet. 2. Return analyzer to monitor mode in Runtime Options window. Vertex M User Manual...
  • Page 130: Remove And Install Pumps

    6.6 Remove and Install Pumps The Vertex M system includes two vacuum pumps. One pump operates while the other is idle. A defective pump may be replaced while the other pump continues to operate. Note: A pump may be replaced only when the system places it in standby. Do not replace an operating pump.
  • Page 131 The pump may be hot. To avoid burns, allow it to cool or wear protective clothing before handling it. Operating the system with the exhaust fitting improperly installed or with one of the pumps removed can result in exposure to toxic gases from remote sampling areas. Vertex M User Manual...
  • Page 132: Install New Pump

    With the system running, verify the leak integrity with a small amount of leak test fluid. 6.6.2 Install New Pump 1. Slide pump assembly into enclosure. 2. Push pump completely into enclosure until the release button locks into the bottom of the enclosure. Vertex M User Manual...
  • Page 133 3. Push on slideplate to re-install on to fitting. 4. Close and fasten the pump enclosure door, install filter, and close lower cabinet door. WARNING Adjacent pump may be hot. To avoid burns, allow to cool or wear protective clothing before handling either pump. Vertex M User Manual...
  • Page 134: Remove And Install Power Supplies

    6.7 Remove and Install Power Supplies 6.7.1 Remove Supply Two replaceable modules provide power for the Vertex M system. A fully populated Vertex M will operate with one supply. The defective supply may be replaced while the system continues to monitor.
  • Page 135: Clean The Touch Screen

    Reference the touch monitor manual for any additional information. 6.9 PLC Module Battery Backup Check Series 1 Vertex M PLC only 1. For status of battery locate indicator as shown in photo. 2. If battery indicator is illuminated the battery has failed. Contact HA Service for corrective action.
  • Page 136: File Maintenance

    Vertex M Technical Manual CD. 6.10.1 Vertex M with Windows 7 1. Touch Project then Stop Project to stop the Vertex M application. To perform this step, an appropriate access rights are required. 2. Close RSView32 by touching File then Exit RSView32...
  • Page 137 3. Make an offline backup copy of the files listed below in Table 6-2. The following files on the Vertex M should be copied to either a backup location on your local area network, removable media, or a CD-R/CD-RW/DVD disk.
  • Page 138: Vertex M With Windows 10

    6.10.2 Vertex M with Windows 10 1. Touch Project then Stop Project to stop the Vertex M application. To perform this step, an appropriate access rights are required. 2. Make an offline backup copy of the files listed below in Table 6-3. The following files on the Vertex M should be copied to either a backup location on your local area network, removable media, or a CD-R/CD-RW/DVD disk.
  • Page 139: Optics Cleaning

    1. Open the Optics Block Gate 2. Remove the Chemcassette™ 3. Open the Vertex M side panel 4. Remove tubing (shown in photo) one at the time and blow out with compressed air. 5. Re-secure side panel and reload Chemcassette™...
  • Page 140: Chapter 7 - Troubleshooting

    TROUBLESHOOTING 7.1 Introduction This chapter helps determine the source of Vertex M system fault or failure and provides a corrective action. There are three general indications that service or maintenance is necessary: General failure to operate properly. See "General System Problems" on the next page for more information.
  • Page 141: General System Problems

    7.2 General System Problems This section applies to problems and solutions that are not reported by the Vertex M software. LCD Display Symptom Problem Recovery Faulty connection at DAq Check connection at SVGA connector on back of DAq Faulty connection at display...
  • Page 142 Check cable connection to Ethernet port Note: Use only the Ethernet connection on the back of the data acquisition Incorrect network connection computer. The Ethernet hub in the Vertex M system is only for Vertex M internal use. Vertex M not appearing on network...
  • Page 143: Maintenance Faults

    7.3 Maintenance Faults A maintenance fault indicates the Vertex M system requires attention but is continuing to monitor. When a maintenance fault occurs, the following actions take place within the Vertex M system: The LED indicator flashes yellow on the analyzer with the fault. See "Manual Analyzer Override" on page  154 for more information.
  • Page 144 Positive pressure between Point pressure above atmospheric pressure while in idle Relieve/reduce pressure sampling location and monitor POS PNT Positive Point Calibration error Contact Honeywell Service Pressure in. Hg PRESSURE Pressure Transducer error Defective sensor Contact Honeywell Service Offset incorrect...
  • Page 145 Replace with supply 0060- POWER SUPPLY Redundant supply not available Failure Fail/missing 0020 1.0: Front block reads <0˚C Cold Environment Relocate Vertex M 2.0: Front block reads >60˚C Electronic problem Replace front optics block Optics BAD OPTICS Sum of error codes 1.0...
  • Page 146 Inadequate cracking pressure Revise installation Sample line leak Observed cracking measured during the Ambient pressure at sampled point differs from Vertex M exhaust by LINE FAIL detected pressure (in Hg) monitoring by the optional Line more than 0.3 in Hg (1 KPa)
  • Page 147 USAGE often that expected Usage advance Old tape Virtual Reference Tape reflectance differs slightly Contact Honeywell Service if VR FAILED Sample Counter Test Failed between windows this occurs frequently Ambient gas concentration *Dependent on the setting in the Configuration menu.
  • Page 148: Instrument Faults

    7.4 Instrument Faults An instrument fault indicates a loss of monitoring on one or more points. When an instrument fault occurs, the following actions take place within the Vertex M system: The LED indicator flashes yellow on the analyzer with the fault. See "Manual Analyzer Override" on page  154 for more information.
  • Page 149 Event History Data Field Problem Possible Cause Recovery Number (Short) Description Check cable Cable disconnected Contact Honeywell service OPTICS Count values from optics reading Optics Failure Counts Optics board defective Contact Honeywell service FAILURE exceeded limits Sensor interface defective Contact Honeywell service...
  • Page 150 Event Event List Desc. Event History Data Field Problem Possible Cause Recovery Number (Short) Description INVALID SUP Supply Pressure Miscalibrated sensor board or defective in. Hg Manifold pressure out of range Replace Analyzer PRES Out Of Range transducer No mains power at points 5-8 Internal thermal fuse opened Check cooling fan, replace thermal fuse No mains power at points 1-4...
  • Page 151 Optics reference OPT REF ADJ Optics block self-adjustment adjustment Error code Cycle power to analyzer. Contact Honeywell service. TIMEOUT timeout timeout Analyzer powered down Check switch on power module OPC To Az ## COM AZ # #...
  • Page 152: Information Events

    7.5 Information Events The Vertex M system enters informational and other non-fault events into the event history database. These do not require any action by the user. Use the event history to check the status of the monitor. Event Type In History...
  • Page 153: Lit Related Informational Events

    Other analyzers in the rack performed a Line Integrity Test but this analyzer was purchased without the option. Note: If additional or replacement Vertex M Analyzers are purchased without specifying the LIT option, no faults will be generated for the lack of the option in the new analyzer. However, an “INFO” message will be...
  • Page 154: Manual Analyzer Override

    7.6 Manual Analyzer Override The Vertex M Analyzer is equipped with a Manual Override (Figure 7-1) button in the event the communications to the Vertex M Data Acquisition (DAq) computer halts. This button activates only when the communications has completely ceased.
  • Page 155 Note: Performing this task will generate Maintenance Fault F111 – DACS COM FAIL 1. Manually release the Analyzer by sliding the Removal key into slot located in the bottom of the Analyzer front (key located in Software Binder shipped with each Vertex M rack) and slide the Analyzer out of the cabinet.
  • Page 156 To Exit Monitor and Open Gate: Press and hold button for 10 seconds and above 5. Close and reattach the Analyzer cover, reconnect the Ethernet communications cable, and then slide the Analyzer into the cabinet. Table 7-1. Analyzer Status LEDs Vertex M User Manual...
  • Page 157: Chapter 8

    CHAPTER APPENDIXES 8.1 Installation Drawings 8.1.1 Introduction Vertex M User Manual...
  • Page 158: Floor Space Requirements

    8.1.2 Floor Space Requirements Vertex M User Manual...
  • Page 159: General Considerations And Cooling

    8.1.3 General Considerations and Cooling Vertex M User Manual...
  • Page 160: General Wiring

    8.1.4 General Wiring Vertex M User Manual...
  • Page 161 Vertex M User Manual...
  • Page 162: 4-Port Manifold Option (P/N: 1295-0322)

    8.1.5 4-Port Manifold Option (P/N: 1295-0322) Vertex M User Manual...
  • Page 163: Floor Mounting Option

    8.1.6 Floor Mounting Option Vertex M User Manual...
  • Page 164 Vertex M User Manual...
  • Page 165: Specifications

    30 days. Filters have an arrow on the side of the filter pointing in the direction of airflow toward the Vertex M system. Replacement of filters is site dependent. Filter A - P/N 780248 Filter B - P/N 1830-0055...
  • Page 166 Hydrogen Selenide Hydrogen Sulfide S LL Hydrogen Sulfide (Low Level) Nitrogen Dioxide Nitrogen Trifluoride COCl Phosgene Phosphine Silane Sulfur Dioxide Tert-butyl Arsine Tert-butyl Phosphine TDMAT Tetrakis (DimethylAmino) Titanium Table 8-1. Sample Filter Requirement Vertex M General Specifications Vertex M User Manual...
  • Page 167: Physical Dimensions

    Weight: ~550 lbs. (249 kg) with 3 analyzers Universal Az: 31 lbs. (14 kg) Pyrolyzer Az: 48 lbs. (22 kg) Empty Vertex M Rack: 457 lbs. (207 kg) Tubing dimensions Sample lines: 1/4 in. (6.35 mm) O.D. x 0.190 in. (4.83 mm) I.D. FEP Teflon or 1/4 in.
  • Page 168: Nominal Transport Times

    8.2.4 Nominal Transport Times The following table shows the time required for samples to move from the sampling point to the Vertex M system for various lengths of sample lines. O.D. 0.25 in. Outside Diameter Tubing I.D. 0.15 in. (Medium Wall) I.D.
  • Page 169: Detectable Gases

    8.3 Detectable Gases Vertex M system Chemcassette™ analyzers are continuous monitoring instruments. The initial analysis period listed in Table C-1 varies based on the programmed alarm levels. This period is valid only after the system pulls a new Chemcassette™ window. Increasing the programmed alarm levels will decrease the initial sample period.
  • Page 170 Default Default Initial Ana- Time to 1 TLV (Alarm @ 2 TLV Concentration 10-ft. Chemcassette Gas Name Alarm Alarm Range Alarm Setting lysis Period Sample Line) Part Number Level 1 Level 2 (second) 0.5-1.9 ppb Arsine XPV- Germane 2-4.9 ppb 5 ppb 0.5 ppb 0.3 ppb...
  • Page 171: Diborane

    Default Default Initial Ana- Time to 1 TLV (Alarm @ 2 TLV Concentration 10-ft. Chemcassette Gas Name Alarm Alarm Range Alarm Setting lysis Period Sample Line) Part Number Level 1 Level 2 (second) 0.030-0.049 ppm Chlorine XPV 0.030 0.010 0.050 0.100 0-1.000 0.050-0.099 ppm...
  • Page 172: Hbr

    Default Default Initial Ana- Time to 1 TLV (Alarm @ 2 TLV Concentration 10-ft. Chemcassette Gas Name Alarm Alarm Range Alarm Setting lysis Period Sample Line) Part Number Level 1 Level 2 (second) 100 - 149ppb Germane XPV- 150 - 199ppb 200 ppb 100 ppb 85 ppb...
  • Page 173: Hcn

    200-399 ppb 400-2000 ppb *Due to U.S. Government regulations, this range may be subject to restrictions requiring special licensing for certain countries outside North America. Contact Honeywell for eligibility information. 0.2-1.4 ppm Hydrogen Iodide (HI) None Est. 0.2 ppm 0.1 ppm 3 ppm...
  • Page 174: H 2 S Ll Hydrogen Sulfide (Low Level)

    Default Default Initial Ana- Time to 1 TLV (Alarm @ 2 TLV Concentration 10-ft. Chemcassette Gas Name Alarm Alarm Range Alarm Setting lysis Period Sample Line) Part Number Level 1 Level 2 (second) 2-9.9 ppm Hydrogen Sulfide 10-19.9 ppm 1 ppm 2 ppm 1 ppm 10 ppm...
  • Page 175: Phosphine

    Default Default Initial Ana- Time to 1 TLV (Alarm @ 2 TLV Concentration 10-ft. Chemcassette Gas Name Alarm Alarm Range Alarm Setting lysis Period Sample Line) Part Number Level 1 Level 2 (second) 10-49 ppb Phosgene XP 50-99 ppb 100 ppb 10 ppb 7 ppb 100 ppb...
  • Page 176 Default Default Initial Ana- Time to 1 TLV (Alarm @ 2 TLV Concentration 10-ft. Chemcassette Gas Name Alarm Alarm Range Alarm Setting lysis Period Sample Line) Part Number Level 1 Level 2 (second) 50-249 ppb Silane XP4 (SiH 250-499 ppb 5 ppm 50 ppb 50 ppb...
  • Page 177: Replacement And Consumable Items

    XP Mineral Acids (Extended Play) 1295-0507 XPV Chlorine-II 1295-0560 XPV Chlorine-3 1295-0565 XPV Germane 1295-0566 XP4-V for Amines/Ammonia 1257-9309 XP4-V for Chlorine 1257-9308 XP4-V for Hydrides 1257-9300 XP4-V for Mineral Acids 1257-9310 XP4-V for Phosgene 1257-9307 Vertex M User Manual...
  • Page 178: Printed Circuit Boards

    1874-0139 Rack Filters Glass Fiber Filter (pump module) 0235-1186 8.4.2 Printed Circuit Boards Pyrolyzers AC Line Filter 1874A0248 Temperature Controller 1295A0466 All Analyzers RFID PCB Assembly 1295A0412 Power Distribution Module Connector PCB 1295A0420 Switch PCB 1295A0422 Vertex M User Manual...
  • Page 179: Components

    Microtube Assembly (one point, for pyrolyzer Pts 1 - 4) 1295A0235 Microtube Assembly (one point, for pyrolyzer Pts 5 - 8) 1295A0236 Check Valve 0235-1157 Pyrolyzer Fan Assembly 1295A0238 Heater Kit (230 Volt, 2 required) 1295K0485 Vertex M User Manual...
  • Page 180 PLC 8 Position Relay Module (Isolated) 0185-0090 PLC 16 Position Relay Module (Non-Isolated) 0185-0053 DH485/RS232 Interface Module 0185-0050 DH485 Link Coupler 0185-0052 Advanced Interface Converter 0185-0051 All other PLC Cards Call Service 2U Computer System Call Service Hard Disk Drive 0185-0078 Vertex M User Manual...
  • Page 181 Hot Swappable 24VDC Power Supply 0060-0020 Pump Modules Pump Assy, 220/230VAC 1290A0059 Pump Assy, 110VAC 1290A0058 Pump Rebuild Kit 0235-0236 Pump Stem and O-Ring 0235-1212 Neoprene Isolation Mount 0950-1061 Thermal Switch (170F) 0170-0082 Fan, 24VDC 0220-0023 Differential Pressure Switch 0050-0039 Vertex M User Manual...
  • Page 182: Series 1 Rack Optional Relay Specifications

    28 VA by the applied DC voltage. For example, 28 VA / 48 VDC = 0.58A. For DC voltage applications less than 14V, the make/break ratings for relay contacts cannot exceed 2A. RTB = Removable Terminal Block. Table 8-3. Relay Contact Ratings for 1746-OW16 Vertex M User Manual...
  • Page 183: Default Relay Assignments

    Introduction By default, the Vertex M reserves the first 2 contacts for faults. Any remaining relay contacts may be associated with points in any analyzer. See "Set Analyzer Window" on page  59 for procedures to associate relay contacts with alarms.
  • Page 184 Terminal Assignment of 1746-OW16 Relay Module Option Part Number: 1290-0076 Individual Relay Card Part Number: 0185-0053 Terminal Assignment of 1746-OX8 Relay Module Option Part Number: 1290-0077 Individual Relay Card Part Number: 0185-0090 Vertex M User Manual...
  • Page 185: Series 3 Rack Optional Relay Specifications

    28VA by the applied DC voltage. For example, 28VA/48V DC = 0.58 Table 8-4: Relay Contact Ratings for 5069-OW16 and OW4I For more information see Allen-Bradley publication 5069-IN018G-EN-P. This can be found at: https://literature.rockwellautomation.com/idc/groups/literature Vertex M User Manual...
  • Page 186 Default Relay Assignments Introduction By default, the Vertex M reserves the first 14 contacts for faults. You may associate any remaining relay contacts with points in any analyzer. See "Set Analyzer Window" on page  59 for procedures to associate relay contacts with alarms.
  • Page 187 2. Strip 12 mm (0.47 in.) of insulation from the wires that you connect to the RTB. 3. Insert the wire into the terminal. 4. Turn the screwdriver to close the terminal on the wire. Torque the screw to 0.4 N•m (3.5 lb•in). Vertex M User Manual...
  • Page 188 For more information about the relay output module, please refer to the Compact 5000 Digital 16-point Relay Output Module Installation Instructions. https://literature.rockwellautomation.com/idc/groups/literature/documents/in/5069-in018_- en-p.pdf Disconnect Wires from the Removable Terminal Block 1. Turn the screwdriver counterclockwise to open the terminal. 2. Remove the wire from the terminal. Vertex M User Manual...
  • Page 189 Note: The 5069-OW4I module does not use SA power. That is, it does not draw current from the SA Power bus. The module passes it through to the next Compact 5000 I/O module in the system. Vertex M User Manual...
  • Page 190: Series 1 Rack Network Interface And Options

    OLE for Process Control (OPC) Interface. See "OLE for Process Control (OPC) Interface" on the facing page for more information. Options: The Vertex M can be fitted with any one of six optional fieldbus interfaces as summarized in table F-1. Protocol Name...
  • Page 191: Ole For Process Control (Opc) Interface

    The OPC drivers are automatically initialized at startup of the application, enabling the Vertex M to act as an OPC Server. The user may connect to the Network interface Cards RJ45 Port at the rear of the data Acquisition System.
  • Page 192 Setting Up an OPC Client Application To request data from Vertex M the OPC Client application will need to include the following information. More information on OPC and client application is available at www.opcfoundation.org. Server Honeywell.VertexOPCServerDA.1 Type Local Enter the Computer Name of the Vertex M Data acquisition PC from Server Computer Settings, Control Panel, System, Network Identification.
  • Page 193 1) Most severe HMI fault 2) If no fault, a blank string String containing the fault number and long fault Vertex M User Manual...
  • Page 194 String representation of timestamp of fault creation associated with Fault\ LFault\Az#. TimeStamp String representation of timestamp of fault creation associated with Fault\ LFault\HMI String representation of timestamp of fault creation associated with Fault\ LFault\PLC Flow Control Vertex M User Manual...
  • Page 195 0 = No fault 1 = Maintenance Fault 2 = Instrument Fault 3 = Both Reports faults from either analyzer or point Unsigned integer Status 0 = No fault Fault 1 = Maintenance Fault 2 = Instrument Fault Vertex M User Manual...
  • Page 196 3-4 - for HA use only bit 5 -- maintenance fault exists bit 6 -- monitor fault exists bit 7 -- for HA use only Maintenance Integer, number of days until CFilter Az#FiltDaysLeft filters should be changed, Vertex M User Manual...
  • Page 197 Table 8-7. Tag Names Note: The Vertex M uses the same OPC server as the Vertex. Thus numerous OPC tags exist for non-existent analyzers. Tags for points greater than 24 or analyzers greater than 3 are not populated and invalid.
  • Page 198: Data Values Common To Fieldbus Networks

    However, indication of faults via the concentration output is disabled by default. It can be turned on in the configuration profile as shown in Figure 8-1. See "Configuration Utility" on page  51 for more information. Vertex M User Manual...
  • Page 199 Each optional fieldbus interface is implemented using commercial communications modules connected to the internal PLC. In the Vertex M the modules function as slaves only, although the hardware may be capable of acting as scanners or masters when installed elsewhere.
  • Page 200: Data Map

    8.7.3 Data Map Five of the optional fieldbusses (all except LonWorks) report the status of the Vertex M in the same 122 word (244 byte) format which is listed in Table F-5. If concentration information is not needed, only the first 41 words must be read. This significantly reduces the consumption of scanner memory.
  • Page 201 Pt 1-2-7 Conc Pt 1-2-8 Conc Pt 1-3-1 Conc Pt 1-3-2 Conc Pt 1-3-3 Conc Pt 1-3-4 Conc Pt 1-3-5 Conc Pt 1-3-6 Conc Pt 1-3-7 Conc Pt 1-3-8 Conc Table 8-10. Fieldbus Data Addresses (Except LonWorks) Vertex M User Manual...
  • Page 202: Profibus Option (P/N 1295-275)

    The data map used by the Profibus interface is shown in Table A-5. The interpretation of the data is in Table F-3 and Table F-4. The network management tool will need a .GSE file to describe every slave on the network. The vertex.gse file can be copied from any Vertex M at c:\hmi\Profibus\vertex.gse.
  • Page 203 Put the SLC 5/03 PLC CPU into run mode by turning the key briefly to RUN then back F1.14. to REM. Verify that the RUN light stays on. Table 8-11. Procedure 1 - Changing Profibus Configuration Vertex M User Manual...
  • Page 204: Devicenet Interface (P/N 1295-0329)

    For more information about ControlNet, see www. controlnet.org. Technical Information on the Allen- Bradley 1747-SCNR ControlNet interface which is used in the Vertex M is available at http://literature. rockwellautomation.com/idc/groups/public/documents/ webassets/browse_ category.hcst. The data map used by the ControlNet interface is shown in Table G-5. The interpretation of the data is in Table G-3 and Table G-4.
  • Page 205: Df1 Interface (P/N 1295-0343)

    Technical information on the ProSoft Technology MVI46-DFCM interface which is used in the Vertex M can be found at http://www.prosoft-technology.com. The baud rate and address of the DF1 interface are set using the configuration profile utility as described in "Configuration Utility" on page  51 Supported speeds range from 1200 to 57,600 baud with a default of 19,200.
  • Page 206: Modbus Plus Interface (P/N 1295-0330)

    This is a purple box mounted below the PLC in the rear of the Vertex M rack. The interface can be operated in either Polled Mode or Bind Mode. A service pin is provided for node identification. LED indicators are provided for Power, Run, System Error, and Configuration Error.
  • Page 207: Modbus/Tcp (P/N 1295-0520)

    Modbus/TCP provides the highly reliable communications like the other fieldbusses over fast, economical Ethernet media. This interface uses ProSoft-Technology MVI46- MNET hardware. In the Vertex M, this interface is always a TCP server, never a client. For more information see www.prosoft-technology.com and www.modbus.org.
  • Page 208 Figure 8-6. Connector Locations Figure 8-7. HyperTerminal Setup for Diagnostic Port Figure 8-8. WATTCP.CFG File Vertex M User Manual...
  • Page 209 A 10/100BaseT Ethernet network cable may now be connected to the top RJ45 jack. Notes: Honeywell strongly recommends that any gas detection network which is connected to the Internet be protected from unauthorized access by a firewall. Honeywell recommends that a gas detection Ethernet network be separated from a general- purpose computer networks by a switch.
  • Page 210 Honeywell, this file will be preloaded. Otherwise, the file provided in C:\hmi\FieldbusFiles must be loaded. The 122 words of Vertex M status may be viewed on the diagnostic port. These are mapped as registers 0 to 121 in the ProSoft database. Instructions on viewing the ProSoft database are in Chapter 6 of the ProSoft User Manual.
  • Page 211: Ethernet/Cip (P/N 1295-0519)

    EtherNet/CIP provides the highly reliable communications like the other fieldbusses over fast, economical Ethernet media. This interface uses ProSoft-Technology MVI46-DFNT hardware. In the Vertex M this interface is always a server, never a client. For more information see www.prosoft-technology.com and www.controlnet.org.
  • Page 212: Vertex M Remote Display Setup

    System Information window. The number of licenses is displayed to the right of Remote Sessions Count. Note: If there are not sufficient licenses or if all licenses are in use, remote access to Vertex M will not be possible. Changing this number is beyond the scope of this procedure; contact a Honeywell representative for more information.
  • Page 213 This number is necessary to establish communication with the rack. The address is available by running a command prompt, the ipconfig command as shown in Figure 8-10. Note that the Vertex M has two Ethernet interfaces: the address of External_Ethernet is the relevant one, as shown in the image below.
  • Page 214 The Java runtime environment must be installed on the desktop to view Niagara Racks. If this is not installed on the PC, it must be installed to use the virtual Vertex M: Click on the Free Java Download button on the webpage (http://www.java.com) and follow the prompts.
  • Page 215 TIP: Adding the Vertex M IP address to the IE Favorites folder saves time. The Vertex M HMI will appear in Internet Explorer as shown in Figure 8-12. The virtual Vertex M interface can be displayed on most screen sizes (1024 X 768 is a minimum, however larger screens are better for this application).
  • Page 216 Previous versions of VertexClient must be removed before installing the newer version (Start/ Control Panel/Add or Remove Programs). CAUTION Protect both the remote station and the network between it and the Vertex M monitor from unauthorized contact. Software changes increase the vulnerability of the Vertex M rack to malicious attack.
  • Page 217 Note: If a DCOM Configuration Warning! window appears, simply click on Yes to record the warning and continue with the procedure. The Component Services tool allows users to set default security for the entire computer and security settings for individual programs. Some programs require special attention. Vertex M User Manual...
  • Page 218 2. Right-click My Computer and select Properties. This opens the My Computer Properties window. 3. Click on the COM Security tab, which contains both Access Permissions and Launch and Activation Permissions. Figure 8-16. COM Security Tab Vertex M User Manual...
  • Page 219 3. In some cases, Windows XP requires that custom permissions be used instead of the equivalent default permissions. 4. Select the program in the Component Services window and right-click to select Properties. 5. Click to select Customize then click Edit as shown on the following page. Vertex M User Manual...
  • Page 220 The recommended workaround is to create a new object Guest. (singular, not Guests plural). This prevents XP from reverting to default. 6. Click Apply after making changes as shown in Figure 8-17. 7. Reboot remote computer after configuration. Figure 8-18. Custom Permissions Vertex M User Manual...
  • Page 221 Allow Guests, CR/CS/C/MC/ NETWORK, Properties LAED Users VertexDCOMUsers Local INTERACTIVE, Launch and SYSTEM Local Activation Administrators, ANONYMOUS LOGON, LAEL Everyone, all Allow INTERACTIVE, NETWORK, SYSTEM Default Authentication None Level Default Properties Default Impersonation Identify Level Vertex M User Manual...
  • Page 222 Table 8-10. Program List for Security Modifications Windows Firewall The Windows Firewall must be turned off: 1. Click on Start in the Windows taskbar 2. Select Settings and then Control Panel from the menu. Figure 8-19. Control Panel Vertex M User Manual...
  • Page 223 Microsoft Windows security requires that matching accounts exist on both the rack and the desktop to provide the necessary credentials. Every Vertex M rack ships with an account named Honeywell and the password vertex One way to provide the matching accounts is to create a similar account on the desktop.
  • Page 224: Demonstrating Remote Opc Communication

    Figure 8-21. Computer Management 8.7.13 Demonstrating Remote OPC Communication This procedure describes how to set up OPC communications from a Vertex M rack and an external Microsoft Windows computer. Some of the required files are provided on the Vertex M Setup Procedure 1.
  • Page 225 10. Install an OPC Client program on the external computer. Screenshots of two examples are included. The Matrikon OPC Explorer is on the Vertex M CD and is pre-installed on the Vertex M rack. It may also be downloaded from www.matrikon.com. Additionally, DAClient from Rockwell Software might be useful and is shown here.
  • Page 226 7. Expand the tree on the left to reveal the Vertex M server as highlighted in purple the Matrikon OPC Explorer window in Figure 8-20. 8. Click on Server then select Add Group. The example name is g1. 9. Select the group as shown highlighted in Figure 8-22.
  • Page 227 1. Start the program. Type in the name of the OPC Server and the OPC server machine as shown in Figure 8-24. Figure 8-25. Enter OPC Server and Machine Name 2. Select a group name such as the default Group1. Vertex M User Manual...
  • Page 228 Figure 8-26. Add Item Form 4. When all desired tags have been added, click OK to close the window 5. Verify that data appears with Current Quality set to Good as shown. Figure 8-27. RSI Data Client Vertex M User Manual...
  • Page 229: Series 3 Plc Network Interface And Options

    All of this is sketched in Figure 8.28. Figure 8.28 The Vertex M presents status as an array of 244 byte as described in Table 8-10 . The values of these data are defined in tables 8-8 and 8-9. Optional gateways translate this data into a variety of fieldbus protocols as listed in Table 8.11.
  • Page 230 Risk of undetected communication loss – it is essential that external automation equipment which communicate with the Vertex M monitor the "Heartbeat" byte 81. (Or the low byte of word 40) A failure of the heartbeat to increment must be treated as a complete loss of gas monitoring.
  • Page 231 Figure 8.30 - Typical Anybus X-gateway, Article Number ABC4*** Article Number AB7* In the Vertex M, the gateway is located at left of the upper PLC DIN rail as highlighted in green in Figure 8.31. Figure 8.31 – Gateway Location in Vertex M The gateways come pre-configured from Honeywell with the full 244-byte memory map.
  • Page 232: Common Considerations Of Anybus Communicator

    This is illustrated in the photograph of the Communicator in Figure 8-33. The configuration port is highlighted in blue. The internal PLC network ports are highlighted in red. And the external fieldbus ports are marked in green. Figure 8.33 Vertex M User Manual...
  • Page 233 An example webpage served by the gateway is shown in Figure 8.35. Figure 8.35 Anybus Communicator Home page The Ethernet/IP section of the gateway must be configured for communication to take place. Gateways purchased from Honeywell will have the internal port preconfigured like this. Vertex M User Manual...
  • Page 234 Figure 8.36 Ethernet/IP Configuration The configuration of the externally facing interface is straightforward and is shown in Figure 8.37. The DHCP checkbox should be unchecked in most installations. Figure 8.37 External Interface Configuration Vertex M User Manual...
  • Page 235 Figure 8.38. Figure 8.38 Adjusting I/O Size The website inside the gateway can monitor the 244-byte Vertex M status array. This can aid debugging. An example of this is shown in Figure 8.39. The byte circled in green indicates that Pt1-1-1 is in level 1 alarm.
  • Page 236: Common Considerations Of Anybus X-Gateways

    1. Connect a USB cable from the PC to the gateway’s USB type B connector on the front. 2. Confirm the gateway is receiving electrical power. 3. Navigate to Online/ Select Connection / Config. 4. Select the current module as shown in green in Figure 8.41. Vertex M User Manual...
  • Page 237 82 bytes. 9. Select Online/Download Configuration. 10. Verify the message box Finished! is displayed. This normally happens in about 30 seconds. 11. Disconnect the USB cable. Vertex M User Manual...
  • Page 238 This section is relevant for setting the internally facing Ethernet interface for X-gateways not purchased from Honeywell and also for the externally facing Ethernet/IP interface. The address may be set via the HMS IPconfig program which can be downloaded from www.anybus.com.
  • Page 239: Profibus - Dp

    Modbus/TCP provides highly reliable communications over economical Ethernet media. For more information, see www.modbus.org. In the Vertex M this interface is always a server, never a client. More information on the ABC4011 Communicator which is used in the Vertex M is available at www.anybus.com.
  • Page 240: Ethernet/Ip

    Ethernet/IP provides highly reliable communications over fast, economical Ethernet media. For more information, see www.odva.org. In the Vertex M this interface is always a server, never a client. Further details on the Anybus X- gateway Article AB7831-F which is used in the Vertex M is available at www.anybus.com.
  • Page 241: Devicenet

    PLCs. DeviceNet network uses a combination of taps and shielded, twisted pair media for device connection. For more information about DeviceNet, see www.odva.org. Technical Information on the AB7833 gateway which is used in the Vertex M is available at www.anybus.com.
  • Page 242 Link OK, online, connected Green (flashing) Online, not connected Critical link failure Red (flashing) Connection timeout Not offline No Power Green Normal Operation Green (flashing) Data Size Error Major fault, unrecoverable Red (flashing) Minor fault, recoverable Table 8.17 Vertex M User Manual...
  • Page 243: Controlnet

    ControlNet offers high-speed, media redundancy and intrinsically safe options. For more information about ControlNet, see www.odva.org. Technical Information on the AB7834-F X- gateway which is used in the Vertex M is available at www.anybus.com. Note that the 244-byte data map is offset by 4 bytes in ControlNet.
  • Page 244: Opc Interface

    The OPC services by FactoryTalk Linx Gateway are automatically initialized at startup of the application, enabling the Vertex M to act as an OPC Server. The user may connect to the Network interface Cards RJ45 Port at the rear of the data Acquisition System. See "Data Acquisition Computer (rear)"...
  • Page 245 FactoryTalk Linx Gateway OPC DA Service For Vertex M system running FactoryTalk View SE, OPC DA is supported with the FactoryTalk Linx Gateway. OPC DA is enabled by default for compatibility with legacy Vertex. The OPC DA service can be configured by opening the FactoryTalk Linx Gateway Configuration application.
  • Page 246 3. As part of the design of OPC DA, the client computer must use the same user and password as the Vertex DAq computer. The Vertex default user is Honeywell Analytics with password vertex. 4. Adjustments to Firewall and DCOM security may be required on the remote OPC DA client computer to allow for communication.
  • Page 247 4. In dialog window, select FactoryTalk Gateway and enter network name or IP address of the Vertex M system in the Node Name field. A connection is established once the network name or IP address of the Vertex server will appear in the application main window header.
  • Page 248 OPC DA connectivity and operability, other OPC DA client application may be used. Demonstration with Matrikon OPC Explorer 1. Install the Matrikon OPC Explorer on remote client, Matrikon OPC Explorer 5.2 was used for this documentation. 2. Start the Matrikon OPC Explorer application. Vertex M User Manual...
  • Page 249 5. If the connection is successful, the OPC DA server and available OPC DA services will be present in the server list panel. Select the FactoryTalk Gateway service on the connected server and select Add Tags button from the OPC Server Connection Options panel to add tags to monitor. Vertex M User Manual...
  • Page 250 Here are some obvious connections items to check: 1. Make sure the User with matching password exist on the OPC DA client and server. 2. Make sure DCOM configuration is configured correctly on both the server and client. Vertex M User Manual...
  • Page 251 Open DCOM configuration, click Windows Start and enter dcomcnfg to run as an Administrator. b. In the Windows Component Services app, right click My Computer, select Properties. Vertex M User Manual...
  • Page 252 For each of the newly added User/Groups, make sure Allow is checked from each permission field, then select OK button when done. f. Add the same Users/Groups for Launch and Activation Permissions; repeat 2c – 2e for Launch and Activation Permissions section in My Computer Properties dialog. Vertex M User Manual...
  • Page 253 In Local User and Groups panel, select Group folder, then right click Distributed COM Users to select Properties from popup menu. c. Select the Add button to enter the user Honeywell Analytics in the Select Users dialog, then select OK button when done.
  • Page 254 FactoryTalk Linx Gateway OPC UA Service For Vertex M system running FactoryTalk View SE, OPC UA is supported with the FactoryTalk Linx Gateway. OPC UA service is disabled by default. To enabled OPC UA service, open FactoryTalk Linx Gateway Configuration to enable OPC UA. OPC UA services have extensive configuration options for security and access control to tags.
  • Page 255 OPC UA client. Certificates may also be rejected or removed from the Trusted Certificates list take away authorized access. Certificates may also be removed from the Rejected Certificates list. Removal of certificates from either list is equivalent to deleting the certificate from the service. Vertex M User Manual...
  • Page 256 FactoryTalk Diagnostic Viewer. This may be useful in identifying issues with OPC communication. OPC UA Client Communication Information needed for connecting to the Vertex M DAq computer OPC UA service for OPC UA client: 1. Discovery URL: opc.tcp://[DAq computer network name or IP address]:4840/FactoryTalkLinxGateway.
  • Page 257 2. Add the Vertex M OPC server information to the OPC UA client. 3. Connect to the Vertex M OPC UA server. If your first attempt to connect to the Vertex M OPC UA server fails, this is likely due to client’s security certificate not trusted.
  • Page 258 4. On the Vertex M, open FactoryTalk Linx Gateway Configuration, if not already open. Select Certificate Management tab, in the Rejected Certificates list, select the certificate that appears to be from the UA client, then press Trust button. After trusting the certificate, the certificate should be added to the Trusted Certificates list.
  • Page 259: Series 1 Rack 4-20Ma Analog Output Option

    8.9 Series 1 Rack 4-20mA Analog Output Option 8.9.1 Overview The application provides an analog output option for the Vertex M by utilizing a second SLC 5/03 PLC rack. This rack communicates with the primary PLC via the DF1 Fieldbus option.
  • Page 260: Operational Description

    PLC memory or communication through the DH485 port. Connecting to this port requires a special adaptor (1747-UIC) and cable (1747-C13). Contact Honeywell Service for assistance. 8.9.4 Operational Description The analog outputs on the secondary PLC will mirror the corresponding Fieldbus concentration register as defined in "Data Values Common to Fieldbus Networks"...
  • Page 261 Vertex M Point Secondary PLC Slot AO Channel on Card 1-2-1 Slot 3 Ch. 0 1-2-2 Slot 3 Ch. 1 1-2-3 Slot 3 Ch. 2 1-2-4 Slot 3 Ch. 3 1-2-5 Slot 4 Ch. 0 1-2-6 Slot 4 Ch. 1...
  • Page 262 Terminal Assignment of Analog Output Module Vertex M User Manual...
  • Page 263: Series 3 Rack 4-20Ma Analog Output Option

    8.10.2 Electrical Isolation The Allen Bradley 5069-OF8 analog output provides sourcing milliampere signals. These are isolated from the Vertex M rack but not from each other. This is shown in the following image. 8.10.3 Hardware Requirements Series 3 Vertex M supports the isolated 4-20mA output using 5069-OF8 analog output modules.
  • Page 264: Configuration Requirements

    2ma fault indicator must be enabled and is off indicator by default. 0 mA on all Communication link failure between PLC and Ethernet Failure channels Adaptor 0 mA on all Including PLC faulted or in program mode Failure channels Vertex M User Manual...
  • Page 265 DIN rail as highlighted in green. The maximum loop resistance is 500 Ω . Channel mapping is defined by the following Table 8-20. Vertex M Point Analog Output Card Output Positive Terminal Number 1-1-1...
  • Page 266 Wiring example Location of Additional Shield Grounding Terminals Vertex M User Manual...
  • Page 267: Line Integrity Test Option

    During the characterization process, the Vertex M records the observed sample pressure. Later LIT tests will issue a fault if the sample pressure increases significantly. Performing the characterization without a leak check invalidates the LIT.
  • Page 268: Quantitative Performance

    The test is not compatible with shared sample lines. The use of external filters is mandatory with Vertex M LIT as it is with all Vertex M points. "Specifications" on page  165 and "Filter Compatibility" on page  165 contain guidance on filter selection.
  • Page 269 The checkmark itself indicates the status of the point- specific checkboxes shown in Figure 8- 48. If the point- specific LIT checkboxes are all set, this checkmark will be set. Figure 8-49. Configuration Editor New Button Vertex M User Manual...
  • Page 270: Lit Access From The Hmi

    The HMI has two buttons in the Authorized Service screen as shown in Figure 8-49. When the associated analyzer is in Monitor mode this will appear as shown in Figure 8-50. Figure 8-50. LIT Access from Authorized Service Mode Figure 8-51. LIT Access from Monitor Mode Vertex M User Manual...
  • Page 271: Lit Related Events

    IDLE to LIT. 8.11.6 LIT Related Events The Vertex M is capable of creating four Maintenance Faults 122 through 125 as shown in "Maintenance Faults" on page  143. If the LIT generates Event 124, care must be taken to inspect the integrity of the sample line tubing along its length as the monitored sample may be taken from the break in the tubing, not from the intended area.
  • Page 272: Warranty Statement

    12 months from the date of first turn-on or 18 months from delivery from Honeywell to the customer, whichever is less. Separate warranty conditions apply to the sensor cartridges limited as indicated below.
  • Page 273: Chemcassette™ Warranty

    Honeywell reserves the right to change this policy at any time. Contact Honeywell for the most current warranty information. 8.12.2 Chemcassette™ Warranty All Chemcassette™ are warranted for a period not to exceed the Chemcassette™ expiration date printed on each package, and tape reel.
  • Page 274: Contact Us

    8.13 Contact Us Find out more sps.honeywell.com Americas Honeywell 405 Barclay Boulevard Lincolnshire, IL 60069 United States Tel: +1 847 955 8200 Toll free: +1 800 538 0363 detectgas@honeywell.com Europe, Middle East, and Africa Life Safety Distribution AG Javastrasse 2...

Table of Contents