Honeywell Vertex M User Manual page 149

Continuous gas monitor
Hide thumbs Also See for Vertex M:
Table of Contents

Advertisement

Event
Event List Desc.
Event History
Number
(Short)
Description
OPTICS
211
Optics Failure
FAILURE
GAS TABLE
212
Gas Table Error
ERROR
HIGH PYRO
High Pyrolyzer
215
FLOW
Flow
OPTICS DRIVE
216
Optics Drive Low
LOW
OPTICS DRIVE
217
Optics Drive High
HI
OPTICS
Optics Reading
218
READING HI
High
DOUBLE PUMP
Double Pump
219
FAIL
Failure
220
LOSS OF FLOW Loss of Flow
AZ SW
Az Software
221
DIAGNOSTIC
Diagnostic
AZ SW
Az Software
223
DIAGNOSTIC
Diagnostic
INVALID PT
Point Pressure
225
PRES
Out Of Range
Data Field
Problem
Count values from optics reading
Counts
exceeded limits
Gas table corrupt or missing
Proportional valve failure
Improperly characterized
flow cc/min.
proportional valve
Static flow reading
Drive value below allowable
Drive counts
threshold
Drive value above allowable
Drive counts
threshold
Drive value above allowable
Optics counts
threshold
Number of
Pump unable to produce 6 in Hg
last pump
vacuum while Fault 112 active.
failed - 1 or 2
Flow less than 180 cc/ min.
On point 1 or 1 and 5 only
cc/min.
Obstruction
Pneumatic integrity
compromised
Pump control conflict between
0.0
one or more analyzers
Pump control conflict between
0.0
one or more analyzers
in. Hg
Sample pressure out of range
Possible Cause
Check cable
Cable disconnected
Contact Honeywell service
Optics board defective
Contact Honeywell service
Sensor interface defective
Contact Honeywell service
Optics LED not properly calibrated
Perform Load CC Operation to recalibrate
No configuration loaded
Reinstall Configuration Profile
Debris in valve
Replace valve
Change in flow system configuration
Perform Auto Balance operation
Transducer failure
Service or replace analyzer
Problem in optics block or sensor
Service or replace analyzer
interface electronics
Problem in optics block or sensor
Service or replace analyzer
interface electronics
Problem in optics block or sensor
Service or replace analyzer
interface electronics
See causes for Fault 112
See recoveries for Fault 112
Circuit breakers are self resetting.
Circuit breaker tripped
Contact Honeywell Service
Perform Auto Balance after any service
Tape tracking problem
Reload Chemcassette
Condensation in system
Purge internal lines
Internal valve filter
Replace filter
Proportional valve
Replace valve
Replace tube
Clogged micro tube
Note: Calibration required
Clogged freon filter (Pyrolyzer)
Replace freon filter
Gate not fully closing
Gate adjustment loose, pivot binding
Optics block loose
Tighten optics block fasteners
Power cycled at random on multiple
Cycle power to resynchronize system
analyzers operating alternate pumps
Power cycled at random on multiple
Cycle power to resynchronize system
analyzers operating alternate pumps
Miscalibrated sensor board or defective
Replace Analyzer
transducer
Positive pressure at sample location
Determine cause of pressure
Recovery