Physical Dimensions - Honeywell Vertex M User Manual

Continuous gas monitor
Hide thumbs Also See for Vertex M:
Table of Contents

Advertisement

8.2.3 Physical Dimensions

Height: 57 in. (144.8 cm)
l
Width: 24 in. (61 cm)
l
Depth: 36 in. (91 cm)
l
Weight: ~550 lbs. (249 kg) with 3 analyzers
l
Universal Az: 31 lbs. (14 kg)
l
Pyrolyzer Az: 48 lbs. (22 kg)
l
Empty Vertex M Rack: 457 lbs. (207 kg)
l
Tubing dimensions
Sample lines: 1/4 in. (6.35 mm)
l
O.D. x 0.190 in. (4.83 mm) I.D. FEP Teflon
l
or 1/4 in. (6.35 mm) O.D. x 0.156 in. (3.9 mm) I.D. FEP Teflon
l
Exhaust line: 1/2 in. (12.7 mm) O.D. x 3/8 in.
l
(9.5 mm) I.D. tubing, 50 ft. (15 m) maximum
l
Output requirements
Optional data output: see "Series 1 Rack Network Interface and Options" on page  190 for more
information.
Optional relay output contacts: Via PLC, normally open (form A). 32 isolated or 64 non-isolated
contacts are user configurable. All relays are undefined by default, and must be configured as
required. See "Series 1 Rack Optional Relay Specifications" on page  182 for more information.
Optional 4-20 mA analog: 0-500 ohms
Installation Category
(Overvoltage category): II (UL 61010B-1)
Temperature: 59°F to 95°F (15°C to 35°C)
Humidity: 20-65% RH
Operating Altitude: –1000 ft. (–305 m) to 6000 ft. (1829 m) above sea level
Operating Voltage: 110 or 230 VAC ± 10% (under load) @ 50/60 Hz; 15 Amps maximum, single
phase. See "Electrical Power" on page  39 for proper specifications.
Vertex M
167
User Manual

Advertisement