5.6.5 Service

The Service Window provides a means to trigger the fault and alarm relays.
Fault Test
Use the Fault Test to verify operation of fault relays.
Note: The fault test simulates an actual fault condition and the Vertex M system activates fault
relays. Notify appropriate personnel that a fault test is planned to be conducted.
To conduct a fault test:
1. Choose Main Screen, Menu, Service and then Fault Test.
2. Choose the analyzer for the fault test.
3. Choose either Instrument Fault or Maintenance Fault.
4. Choose Create Fault Now to activate the relay or relays connected to the selected analyzer
and a fault will appear in the event list.
5. To continue testing, repeat steps 2 through 4.
See "Event List" on page  100 to clear events.
Vertex M
111
User Manual