Setting Up an OPC Client Application
To request data from Vertex M the OPC Client application will need to include the following
information. More information on OPC and client application is available at
www.opcfoundation.org.
Server
Type
Server Computer
Name or Address
Access Path
Update Rate
Item
Folder
Count
Alarm
72
AlarmLatched
72
CommStatus
9
Diagnostics
9
PyroTemp
72
Vertex M
Honeywell.VertexOPCServerDA.1
Local
Enter the Computer Name of the Vertex M Data acquisition PC from
Settings, Control Panel, System, Network Identification.
Typically: Vertex M_291-####, with the # being the 4 digit serial number.
Project name. This can be left blank. If name is specified, use the name of
the open project.
A period in seconds.
Tag name. Whether to specify an item depends on the client application.
Table 8-6 Additional OPC Client Information
Recommended
Tag
for most HMI
systems
Y
Tag Name
Value Interpretation / Descrip-
Format
Integer, 1 byte long Status
0 = No alarm
##_1
1 = Warning
2 = Alarm 1
3 = Alarm 2 Always nonlatching
Integer, 1 byte long Status
0 = No alarm
1 = Warning
Ch##
2 = Alarm 1
3 = Alarm 2
Latching if and only if rack
configured for latching alarms.
String, N/A (unconfigured), NO
CmmFlag#
(not communicating), or YES
(correct)
Float32, Equivalent to
Chemcassette Tape Left as
Az#CCDaysLeft
displayed in Diagnostic /
Analyzer Info
Float32 Equivalent to Pyrolyzer
##
Temperature as displayed in