Honeywell Vertex M User Manual page 145

Continuous gas monitor
Hide thumbs Also See for Vertex M:
Table of Contents

Advertisement

Event
Event List Desc.
Event History
Number
(Short)
Description
Optics Drive
113
OPTICS MAINT
High-Cleaning
Req
Excessive Point
114
POINT VAC ERR
Vacuum
AZ SW
Az Software
115
DIAGNOSTIC
Diagnostic
PUMP OVER-
Pump Over
116
TEMP
Temperature
PUMP EXHAUST
High Exhaust
117
HI
Pressure
Power Supply
118
POWER SUPPLY
Fail/missing
Optics
BAD OPTICS
119
Temperature Out
TEMP
Of Range
Data Field
Problem
LED drive required to set zero
too high
Drive
Tape leader installed
improperly
Sample vacuum is greater than
in. Hg
5 in. Hg vacuum for longer
than 5 seconds
Analyzer failed to enter monitor
0.0
after pump swap
Pump module Temperature
0.0
exceeded
Exhaust Manifold pressure
0.0
greater than 9 in. Hg (4.4 PSI)
0.0
Redundant supply not available Failure
1.0: Front block reads <0˚C
2.0: Front block reads >60˚C
Sum of error codes 1.0
to 64.0
4.0: Front block reads 45-60˚C
8.0: Rear block reads <0˚
Possible Cause
Check valve on inactive pump leaking
Leader not centered (Pts 1 or 5 error)
Autozero performed while tape on light or dark gray portion of leader
Dirt in optics
End of line filter clogged
Sample line kinked
Maximum line length exceeded on a shared sample line
Crimped tube in cable carrier
Control related fault is issued because of an improper control response
from another analyzer
Clogged Filter
Fan Failure
Line voltage less than 208 VAC
Kinked exhaust
Exhaust tubing length exceeds 50 ft. (15 m.)
Cold Environment
Electronic problem
Cooling air failure
Hot environment
Cold environment
Recovery
Analyzer)
Replace check valve
Contact Honeywell Service
Clean optics
Contact Honeywell Service
Reload Chemcassette and
recalibrate using leader
Reload Chemcassette and
recalibrate using leader
Clean optics block
Replace clogged end of line
filter
Isolate by disconnecting
possible crimped sample line
Isolate shared sample line
Review limits number of
analyzers, tubing ID vs. Length
(See "Nominal Transport Times"
on page 168 for more
information.)
Identify crimps in tubing
harness by checking pressure
with analyzer closed vs. open
Verify all analyzers enter
monitor mode successfully
Replace air filter
Check fans in pump module
Verify main line voltage
Check exhaust tubing for kinks
or restrictions
Reroute to reduce length or
change to a larger diameter
sample line
Replace with supply 0060-
0020
Relocate Vertex M
Replace front optics block
Change air filter, replace fan
Relocate Vertex M
Relocate Vertex M