Chemcassette™ Detection System - Honeywell Vertex M User Manual

Continuous gas monitor
Hide thumbs Also See for Vertex M:
Table of Contents

Advertisement

The high speed of transport flow allows rapid monitoring and response time when using long
lines from monitored locations to the Vertex M system. A small portion of the transport flow
(sample flow) is analyzed to determine concentration levels.
The complete sampling and monitoring system consists of the following components:
Sample lines to all monitored locations
l
Flow connections through quick-connect ports in bulkheads on top of unit
l
Moving cable and connectors
l
Vacuum pumps
l
Analyzers incorporating manifolds, Chemcassette™ and filters
l
Flow controlling proportional valve
l
Top exhaust port
l
There are 24 inlets, one for each monitored location. One exhaust port is also located on top of
the Vertex M cabinet.
2.7 Chemcassette™ Detection System
The Chemcassette™ Analyzer module is a self- contained, microprocessor controlled analyzer
that occupies one slot in a Vertex M. Sample lines and the vacuum source are connected to the
Chemcassette™ via a single 10-tube connector.
The system powers up in the same state as when powered down. Data is stored in the module's
memory until the data acquisition computer retrieves it.
The Vertex M Analyzer modules use the Honeywell Chemcassette™ optical detection system.
Analyzer modules sample and detect a specific gas or family of gases.
Each eight-point Analyzer module:
l
Manages Chemcassette™ tape transport
l
Provides optical detection of stain
l
Directs sample flow through the Chemcassette™
l
Stores data for retrieval by the data acquisition computer
l
Components of the detection system include:
Chemcassette™ detection tape
l
Optics and electronics for the detection system
l
Chemcassette™ tape transport mechanism
l
Self adjusting proportional valves
l
Vertex M
26
User Manual

Advertisement