3.4.3 Installing Sample Line Particulate Filters

Attach a sample line filter to the sampling end of the line for all locations.
Keep in mind that excess amounts of dirt in the filters reduces the sample flow, raises
sample vacuum, and may affect concentration readings of the analyzer.
See "Specifications" on page  165 to determine the proper filter type to use with each target gas.

3.5 Installing Pump Exhaust Line

This section describes exhaust connections and installation. The Vertex M system is equipped
with a vacuum pump that is located in the bottom of the Vertex M system cabinet. The pump
exhaust line connects to the manufacturing facility central toxic exhaust system.

3.5.1 Exhaust Line Installation Requirements

Follow the general requirements listed below when installing exhaust lines.
The length of the line should not exceed 50 ft. (15 m). If longer distances are required, contact
Honeywell.
Do not crimp exhaust lines, or place them in an area where weight could collapse the tubing, or
bend them to less than a 12 in. (30.5 cm) radius.
Where possible, leave as many bends exposed for periodic visual inspection of the line for
kinked or damaged tubing.
Varying exhaust pressure can induce pump failure or flow faults.