7. Reach behind analyzer and remove four electrical cables.
8. Unlock and remove tubing harness.
9. Unlatch the slides.
10. Support the analyzer and remove it from the slides.

6.4.3 Install Analyzers

1. Reach into the slot and position cables out of the way.
2. Place analyzer on slides. Push in until locks engage.
3. Slide Analyzer out until fully extended.
4. Reach behind the analyzer, connect, and lock the tubing harness.
Note: When reconnecting the analyzer, connect the Analyzer Communications cable to the
analyzer first.
5. Connect the Analyzer Communications, 24V Power Supply, Multifunction Connector and
ChemCam cables.
6. Carefully push analyzer partially into the cabinet and then pull out to verify all cables move
freely and the slides lock. Repeat in and out action to loosen slide. Push analyzer into
cabinet.
Return to Service
1. Open the front door and turn analyzer power switch (and pyrolyzer power switch if
necessary) on.
2. Re-install Configuration Profile.
3. Install Chemcassette™.
4. Return analyzer to monitor mode in Runtime Options Menu.