Lit Related Informational Events - Honeywell Vertex M User Manual

Continuous gas monitor
Hide thumbs Also See for Vertex M:
Table of Contents

Advertisement

Event Type In History
INFO
Run-Time Alarm Disable
INFO
Run-Time Alarm Disable Cancel
INFO
Run-Time Point Disable
INFO
Run-Time Point Disable Cancel
INFO
Start Monitor
INFO
Stop Monitor
INFO
User Requested Pump Switch
INFO
OPC Driver Started
INFO
OPC Driver Shut Down
INFO
CPLD Programmed Successfully
INFO
CPLD Program Failed
INFO
RFID Read Skipped
OPC
No Ack Echo From Az n-n
OPC
No Cmnd Socket Sent To Az n-n
OPC
OPC To Az n-n Communication Broken
OUT OF ALARM
Various
PLC CONF INSTALL
Various
RESET ALM n
Various
RESET ALM SIMULATE
Various
RESET INST FAULT
Various
RESET MAINT FAULT
Various
USER RESET
Various
INFO
Flow Corrected

7.5.1 LIT Related Informational Events

Event Type in History
Event History Message
INFO
Command - Perform LIT Char
INFO
LIT Characterization Performed
INFO
Command - Unscheduled LIT
INFO
Line Integrity Test Performed
INFO
Analyzer Lacks LIT
Note: If additional or replacement Vertex M Analyzers are purchased without specifying the LIT option, no faults will be generated for the lack of the option in the new analyzer. However, an "INFO" message will be
logged in the Event History each time the LIT option is invoked. Existing analyzers configured for LIT will be unaffected.
Event History Message
User initiated action
User initiated action
User initiated action
User initiated action
User or power restored
User or instrument fault
User initiated action
The DAq booted
User initiated action
A new program has been loaded into the sensor board CPLD chip
An attempt to load a new program into the sensor board CPLD chip failed
User loaded new Chemcassette but skipped reading the RFID tag by pressing <next>
Communications problem to analyzer
Communications problem to analyzer
Communications problem to analyzer
Non-latching alarm level decreased because of decrease in gas concentration
Download PLC Configuration was started
Latching alarm level decreased because user pressed RESET CURRENT or RESET ALL
Analyzer acknowledges command to reset an artificial alarm OR fault
Analyzer acknowledges command to reset an instrument fault
Analyzer acknowledges command to reset a maintenance fault
A user pressed RESET CURRENT or RESET ALL
Flow on one window of tape improper
User pressed START LIT CHAR button
Analyzer completed LIT characterization. If data field non-zero, this indicates the observed cracking pressure in inches Hg.
User pressed START LIT TEST
Analyzer completed an integrity test. The data field contains the observed cracking pressure in inches Hg.
Other analyzers in the rack performed a Line Integrity Test but this analyzer was purchased without the option.
Possible Cause
Possible Cause