7.4 Instrument Faults

An instrument fault indicates a loss of monitoring on one or more points. When an instrument fault occurs, the following actions take place within the Vertex M system:
The LED indicator flashes yellow on the analyzer with the fault. See "Manual Analyzer Override" on page  154 for more information.
l
A yellow fault indicator appears on the main screen
l
The event list and event history are updated
l
Maintenance Fault relays associated with this analyzer will activate
l
Event
Event List Desc.
Event History
Number
(Short)
Description
HIGH
202
High Background
BACKGROUND
End of
203
END OF CC
Chemcassette
FAIL WR HW
Failure Writing
205
CONF
Hardware Config
FAIL RD HW
Failure Reading
206
CONF
Hardware Config
207
PYRO FAILURE
Pyrolyzer Failure
PYRO
Skin Over-
208
OVERTEMP
Temperature
Gate Motor
209
GATE TIMEOUT
Timeout
210
OPTICS NOISE
Optics Noise
Data Field
Problem
Bad Chemcassette
Optical signal
Chemcassette tracking/
counts
installation error
Optics block dirty
Insufficient Chemcassette
0.0
motion detected
Failure writing hardware
configuration
Failure reading hardware
configuration
Internal Temp
Pyrolyzer failure
1 or 2
Pyrolyzer skin temperature
Pyrolyzer
greater than 105°C
number
0.0
Gate motor timeout 7 seconds
Counts
Optics Noise
Possible Cause
Chemcassette discolored
Check optics values in event history
Moisture on Chemcassette
Check optics values for last Chemcassette load in event history
Check guide position
Errors occurring on points 1 and 5 or 4
and 8 only indicate a tracking problem
Reload Chemcassette using maintenance analyzer operation
Dust
Clean optics
End of Chemcassette
Replace Chemcassette
Rethread Chemcassette (See "Change Chemcassette™ Tape" on
Chemcassette broken
page 124)
Faulty tape encoder
Service analyzer
Advance motor failure
Service analyzer
Gate opening insufficient
Service analyzer
Non-volatile memory failure in analyzer
Service or replace analyzer
CPU
Non-volatile memory failure in analyzer
Service or replace analyzer
CPU
Heater element
Fuse failure
Service or replace analyzer
Thermocouple Failure
Fan failure in analyzer
Replace fan
Sensor failure
Replace sensor
Gate position sensor not activated before
Check motor operation using Maintenance/ Analyzer Operations/
timeout
Open Gate or Close Gate
Motor does not operate
Check motor connections to sensor interface PCB in analyzer
Check sensor connection on PCB
Bad sensor or cable
Contact Honeywell service
Poor grounding
Contact Honeywell service
Optics block cover loose
Retighten or reinstall as required
Recovery