Honeywell Vertex M User Manual page 151

Continuous gas monitor
Hide thumbs Also See for Vertex M:
Table of Contents

Advertisement

Event
Event List Desc.
Event History
Number
(Short)
Description
AZ SW VER
Analyzer software
235
WRONG
version mismatch
AZ VACUUM
Inadequate
237
LOW
Analyzer Vacuum
Optics block
OP SW
238
software
MISMATCH
mismatch.
Optics block
OPTICS COMM
239
communication
FAIL
failure
OPTICS REF
Optics reference
240
FAIL
limit exceeded
Optics reference
OPT REF ADJ
241
adjustment
TIMEOUT
timeout
OPC To Az ##
COM AZ # #
270
Communications
BROKEN
Broken
PLC-COM
301
PLC-Com Broken
BROKEN
AZ-BAD PROF
AZ has wrong
302
ID
profile loaded
PLC-BAD PROF
PLC Has Wrong
303
ID
Profile loaded
PLC SYSTEM
304
PLC System Fault
FAULT
DATA LOG
Unable To Log
305
FAILED
Conc Data
CMD ERR AZ #-
413
Cmd Error Az #-#
#
Data Field
Problem
0
Software mismatch
Observed
Individual analyzer observes
vacuum in in
inadequate vacuum, stops
Hg
monitoring
The two optics blocks are not
0
running the same software
version.
Analyzer is not able to
Error code
communicate with the optics
block.
The signal detected at the
Reference
reference photodiode is out of
value signal
range.
Optics block self-adjustment
Error code
timeout
Analyzer not responding to OPC
driver polling request.
PLC fails to respond to DAq
communication
Profile ID in analyzer does not
match current configuration in
use
Profile ID in PLC does not match
current configuration in use
A fault has occurred on the PLC
PLC Error
that prevents the PLC program
Code
from running
Failure to store concentration
data in the database
Analyzer not responding DAq
Communications
Possible Cause
Reload program to all analyzers
Contact Honeywell service
Fault 219 has triggered.
Recover from Fault 219, then reset Fault 237
Circular tubing harness disconnected
Connect circular tubing harness
Correct tubing dress in cable carrier
Tubing pinched
Contact Honeywell Service
Program update done incorrectly.
Reload program to both optics blocks.
Cable disconnected.
Check the 14 pin cable connection on the sensor board.
Software anomaly.
Cycle power to analyzer.
Defective optics block.
Contact Honeywell service.
Degraded optics block.
Contact Honeywell service.
Cycle power to analyzer. Contact Honeywell service.
Analyzer powered down
Check switch on power module
Check Ethernet connection at the back of the analyzer.
Ethernet cable not connected
Move cable to different port on Ethernet hub.
Series 1 Vertex M PLC
Check serial communications on DAq COM2
Check serial cable on PLC 1747-KE (See "Data Acquisition Computer
Cable connection problem
(rear)" on page 19 for more information.) (See "Main PLC" on page 22
for more information.)
Series 3 Vertex M PLC
Check Ethernet connection between PLC and network switch
Loss of PLC power
Check power connections for PLC
Configuration not properly loaded
Reinstall current Configuration Profile
Configuration not properly loaded
Reinstall current Configuration Profile
Faulty or missing PLC module
Contact Honeywell Service
PLC is not properly configured
Local database file has reached size limit
Enable automatic database purge or adjust database purge settings
(1GB maximum size for Access
to purge old data
database).
Cable connections
Check Ethernet connections
Analyzer missing or powered down
Remove analyzer from Configuration Profile
Recovery