Maximum flow rate
Fraction collection can be performed at different maximum flow rates depending on
what type of deep well plates that are used. The table below lists the maximum flow
rates for the different plate types.
Plate type
96 deep well plate
48 deep well plate
24 deep well plate
ÄKTA avant User Manual 29035184 AE
3.10.3 Cassettes, cassette tray and racks
Maximum flow rate (ml/min)
10
15
25
3 Standard instrument modules
3.10 Built-in fraction collector