Conductivity monitor
Item
Number of monitors
Conductivity reading range
Accuracy
Operating pressure
Flow cell volume
Temperature monitor integrated
in Conductivity monitor
Item
Reading range
Accuracy
pH monitor
Parameter
pH reading range
Accuracy
Operating pressure
Flow cell volume
Outlet valve fractionation
Parameter
Number of outlets
ÄKTA avant User Manual 29035184 AE
Description
Up to 2
0.01 to 999.99 mS/cm
±0.01 mS/cm or ±2%, whichever is greater,
(within 0.3 to 300 mS/cm)
0 to 5 MPa (0 to 50 bar)
22 µl
Description
0°C to 99°C
±1.5°C within 4°C to 45°C
Data
0 to 14
± 0.1 pH unit
(within pH 2 to 12, temp. within 3°C from calibration temp.)
0 to 0.5 MPa (0 to 5 bar)
ÄKTA avant 25: 76 µl
ÄKTA avant 150: 129 µl
Data
10, expandable to 32