Reference tubing
The table below shows the label, diameter, and standard length of the reference capillary.
The capillary is used during the System performance tests. The capillary is not mounted
at delivery.
Label
Description
Ref 1
Reference capillary
Outlet tubing
The table below shows the labels, diameters, and standard lengths of the outlet tubing.
The tubing is not mounted on delivery.
Label
Description
Out1 -
Outlets from Outlet valve
Out32
V9-O or V9H-O
Waste tubing
The table below shows the labels, diameters, and standard lengths of the waste tubing.
The waste tubing is mounted on delivery.
Label
Description
W1
System pump waste
W2
Sample pump waste
W3
pH valve waste
W
System waste
N/A
Top tray waste
N/A
Fraction collector F9-C
waste
ÄKTA avant User Manual 29035184 AE
Tubing
PEEK, o.d. 1/16", i.d. 0.25 mm
Tubing
V9-O: ETFE, o.d. 1/16", i.d. 1.0 mm
V9H-O: FEP o.d.1/8", i.d. 1.6 mm
Tubing
ETFE, o.d. 1/16", i.d. 1.0 mm
ETFE, o.d. 1/16", i.d. 1.0 mm
ETFE, o.d. 1/16", i.d. 1.0 mm
ETFE, o.d. 1/16", i.d. 1.0 mm
Silicone, o.d. 12 mm, i.d. 8 mm
Silicone, o.d. 12 mm, i.d. 8 mm
9 Reference information
9.3 Tubing and connectors
Length
Volume (μl)
(mm)
400
20
Length
Volume
(mm)
(ml)
1500
1.2
3.0
Length
Volume
(mm)
(ml)
1500
1.2
1500
1.2
1500
1.2
1500
1.2
1500
80
1350
70
439