8 Troubleshooting
8.3 Troubleshooting: Monitors
Problem
The internal valve tem-
perature is too high
376
Possible cause and action
The air intake on the rear or on the left side of the
•
instrument is covered
Make sure that none of the air intakes on the instru-
ment are covered.
Hot surroundings
•
Decrease the room temperature. Maximum operating
temperature is 35°C.
Hardware error
•
Switch off the instrument and wait until the tempera-
ture has decreased. Restart the instrument. If this error
is recurrent, generate a System error report and con-
tact Service.
ÄKTA avant User Manual 29035184 AE