8 Troubleshooting
8.3 Troubleshooting: Monitors
Problem
The internal valve tem-
perature is too high
376
Possible cause and action
The air intake on the rear or on the left side of the
instrument is covered
Make sure that none of the air intakes on the instru-
ment are covered.
Hot surroundings
Decrease the room temperature. Maximum operating
temperature is 35°C.
Hardware error
Switch off the instrument and wait until the tempera-
ture has decreased. Restart the instrument. If this error
is recurrent, generate a System error report and con-
tact Service.
ÄKTA avant User Manual 29035184 AE