9 Reference information
9.8 Using Manual instructions
9.8.2 Manual instructions - Flow path
Instruction name
Versatile valve to Versa-
tile valve 4
Valve X1
Valve X2
Injection mark
480
Description
Turns the numbered Versatile valve to the selected po-
sition.
Note:
Four sets of positions are available. In positions 1-3 and
2-4 only a single flow channel can be used. In positions
1-4 & 2-3 and 1-2 & 3-4 the flow can be directed through
two channels simultaneously.
Turns the extra inlet valve to the selected position. The
extra valve is a basic 8 port valve, without air sensor, to
be used for general applications.
Turns the extra inlet valve to the selected position. The
extra valve is a basic 8 port valve, without air sensor, to
be used for general applications.
Sets an injection mark in the chromatogram at the point
where this instruction is executed.
Note:
The instruction is useful when the sample is loaded onto
the column by the system pump.
ÄKTA avant User Manual 29035184 AE