Step
3
Fill the Superloop using a syringe
Follow the instruction below to fill the Superloop using a syringe.
Step
1
ÄKTA avant User Manual 29035184 AE
Action
Connect a piece of tubing from the bottom of the Superloop to:
port LoopF on the Injection valve
•
or
the F port corresponding to the connected E port, e.g., 1F, on the Loop
•
valve
Action
Check if the system is in state Ready.
If yes: The Injection valve is in position Manual Load per default. Continue
•
to step 3.
If no: Continue to step 2 to position the valve.
•
5.7 Sample application
5.7.2 Sample application using a Superloop
5 Operation
195