8 Troubleshooting
8.6 Troubleshooting: Pumps
Problem
Leakage around a con-
nector
Erratic pump pressure
The pump is not found
by the instrument
The internal tempera-
ture of the pump is too
high
398
Possible cause and action
Leaking connection and/or crystallized material
around a connector
Unscrew the connector and check if it is worn or incor-
rectly fitted. If so, replace the connector. Gently finger
tighten the connector.
Air trapped in the pump heads
Partially blocked solvent filters
Leaking connections
Piston seal leakage
Check valve malfunction
Piston damaged
See Examples of pump pressure curves, on page 400
for examples of pump pressure curves.
The cable between the pump and the ICU is not
connected.
Contact Service.
A fuse in the instrument ICU is broken.
Contact Service.
The air intake on the rear or on the left side of the
instrument is covered
Make sure that none of the air intakes on the instru-
ment are covered.
Hot surroundings
Decrease the room temperature. Maximum operating
temperature is 35°C.
Hardware error
Switch off the instrument and wait until the tempera-
ture has decreased. Restart the instrument. If this error
is recurrent, generate a System error report and con-
tact Service.
ÄKTA avant User Manual 29035184 AE