8 Troubleshooting
8.3 Troubleshooting: Monitors
Problem
Drift of pH signal when
the pH electrode has
been removed from
storage solution
Temperature reading
error
It is not possible to in-
ject calibration solu-
tion
Alarm in UNICORN:
(Alarm) The pH cell can
only be run at pres-
sures below 0.8 MPa.
Please check the tubing
or lower the flow
through the pH cell.
Note: The pressure limit
0.8 MPa is for the post
column pressure.
378
Possible cause and action
Decreasing salt concentration in the electrode
membrane due to osmosis to buffer
Regenerate the pH electrode. Place the electrode in
deionized water for 30 minutes followed by 30 minutes
in a buffer with pH 4.
The temperature compensation of the pH monitor
is turned off
Contact Service.
Waste tubing is twisted or blocked
Untwist the tubing.
Perform System CIP to clean waste tubing. See Sec-
tion 7.6.2 Perform System CIP, on page 289.
Change the tubing.
High pressure in the pH cell
Decrease the flow rate.
By-pass the pH electrode (see Ports and flow paths of
the pH valve, on page 81) and measure pH in fractions
manually.
If ÄKTA avant 25 is used: Replace tubing between
Column valve and Outlet valve by tubing with i.d. 0.75
mm (tubing: 6, 7, 8, 9, 1R and 2R).
ÄKTA avant User Manual 29035184 AE