9 Reference information
9.15 Check and change the Node ID of a module
Check and change the Node ID
The Node ID of a module is set by the positions of an arrow of two rotating switches at
the back of the module. Follow the instructions to check or change the Node ID.
Step
1
2
3
4
5
518
Action
If applicable, remove the module according to the instruction in Install a
module in the instrument, on page 103.
The Node ID is set by the positions of an arrow of two rotating switches at
the back of the module.
The first rotating switch, labeled A sets the tens.
The second switch, labeled B sets the units.
For example for Node ID 13, the A switch is set to 1 and the B switch to
3.
Check the Node ID and compare it with the listed Node IDs in the tables
above.
To change the Node ID, use a screwdriver to set the arrows of the switches
to the desired number.
Re-install the module in the instrument, if applicable.
A
B
ÄKTA avant User Manual 29035184 AE