Constraints on optional modules
The table below indicates usage constraints for the different external modules.
External module
I/O-box E9
I/O-box E9, 2nd
Fraction collector 2
External air sensor
Optional module mounted in Ex-
tension box
Note:
Edit System Properties
When a new module has been installed, the system properties have to be updated in
UNICORN. The system will restart automatically when the configuration has been changed
in the System Properties dialog box and then the system can be reconnected.
There are four main types of modules (named components in UNICORN) to select from:
More than one component of the same type is only available if the first has been selected.
The selection made is reflected in which instructions and phase properties that are
available.
The following instruction gives a general description of how to update the system
properties in UNICORN.
ÄKTA avant User Manual 29035184 AE
To optimize signal quality, the total cable length connecting all external modules
to the ÄKTA avant instrument should not exceed 10 m.
Valves and pumps
Monitors and sensors
Fraction collectors
Other (e.g., I/O-box)
4 Optional instrument modules
4.2 Installation of optional modules
Constraints
I/O-box E9 has no constraints.
I/O-box E9, 2nd requires I/O-box E9.
Fraction collector 2 requires Fraction collector
and can not be installed at the same time as
Outlet valve 2nd or 3rd.
External air sensor has no constraints
See respective section for each module.