GE AKTA avant User Manual page 119

Hide thumbs Also See for AKTA avant:
Table of Contents

Advertisement

System properties
Follow the instruction to update the system properties.
Step
1
2
3
System settings
If the valve is placed in the flow path between the UV monitor and the outlet valve, the
delay volume must be set. See Check or set delay volumes, on page 153.
Note:
ÄKTA avant User Manual 29035184 AE
Action
Open the system properties Edit dialog box.
In the Component types list, click Valves and pumps.
In the Component selection list, select the check boxes corresponding to
the installed modules. Then click OK.
It is recommended not to alter the default values for restrictor and pH cell delay
volumes when standard modules and standard tubing for flow restrictor are
used.
4 Optional instrument modules
4.5 Extra Outlet valves
119