9.8 Using Manual instructions
9.8.3 Manual instructions - Monitors
Instruction name
Relative scale cond 2nd
482
Description
Facilitates monitoring using a second conductivity
monitor, of a gradient, for which the user sets the con-
ductivity values for 0% and 100%. The Relative scale
cond instruction can be set in ascending manner (0%
for low and 100% for high conductivity) or in descending
manner (0% for high and 100% for low conductivity).
Note:
The Relative scale cond instruction in descending man-
ner is especially useful for conductivity visualization in
RPC and HIC, where the conductivity curve is reversed
compared to the concentration curve (i.e., high conduc-
tivity at 0% B and low conductivity at 100% B).