Pressure monitors
Problem
Pressure offset
Excessively high pres-
sure values
The pressure monitors
are not found by the
instrument
The internal tempera-
ture of the pressure
monitor is too high
ÄKTA avant User Manual 29035184 AE
Possible cause and action
The monitors have lost their calibration
Calibrate the pressure monitors. See Section 7.7.2
Calibrate the pressure monitors, on page 323.
The temperature has changed
Wait until the temperature has stabilized and calibrate
the pressure monitors.
Incorrect installation of extra Column valve
Check installation positions and tubing connections.
Refer to the installation instructions delivered together
with the extra valve.
Unclean inline filter in the Mixer
Replace the inline filter in the Mixer. See Section 7.3.2
Replace the inline filter, on page 275.
Solids in the flow path
To use the predefined method System CIP to clean the
flow path, see .
To clean the column, see Section 7.6.2 Perform System
CIP, on page 289. If persistent, replace the column.
The cable between the Pressure monitors and the
ICU is not connected.
Remove the monitor and make sure that the cable is
connected.
The air intake on the rear or on the left side of the
instrument is covered
Make sure that none of the air intakes on the instru-
ment are covered.
Hot surroundings
Decrease the room temperature. Maximum operating
temperature is 35°C.
Hardware error
Switch off the instrument and wait until the tempera-
ture has decreased. Restart the instrument. If this error
is recurrent, generate a System error report and con-
tact Service.