5 Operation
5.9 Create a method and perform a run
5.9.2 Perform a run
Perform a manual run
Manual runs can be convenient for procedures such as filling tubing with buffer or
packing a column with media.
Step
1
2
3
4
5
Tip:
232
Action
On the Manual menu, click Execute Manual Instructions.
Select instruction group and instruction.
Select or enter parameter values.
Click Insert to have several instructions performed at the same breakpoint.
Note:
Manual runs are only stored temporarily. However, you can choose to store
them permanently in a selected directory. To save results in a chosen directory,
click Browse before the run is started.
Note:
If a method run is started during a manual run, the results from the manual
run are not stored.
To perform the instructions, click Execute.
Manual runs can also be controlled from the Process Picture.