8 Troubleshooting
8.3 Troubleshooting: Monitors
Conductivity monitor
Problem
The Conductivity moni-
tor is not found by the
instrument
The cell constant mea-
surement has been
aborted
Unstable conductivity
Temperature out of
range for calibration
372
Possible cause and action
The Conductivity monitor cable has come loose
Switch off the instrument.
Reconnect the Conductivity monitor cable (the black
cable in the illustration below).
Restart the instrument.
Internal errors
See error log. Restart instrument and retry. If this
problem recurs, generate a System error report and
contact Service.
Air in the Conductivity flow cell
Flush the Conductivity flow cell with water.
Solids in the Conductivity flow cell
Clean the Conductivity cell. See Section 7.6.12 Clean
the Conductivity flow cell, on page 318.
This error can only occur when the temperature
compensation is turned on. The error will occur
when the temperature is outside the range 2°C to
40°C.
Make sure the temperature of the calibration solution
is within 2°C and 40°C.
ÄKTA avant User Manual 29035184 AE