GE AKTA avant User Manual page 427

Hide thumbs Also See for AKTA avant:
Table of Contents

Advertisement

Number of inlets
Parameter
Inlet A
Inlet B
Sample inlet
Quaternary inlet
Pressure monitors
Parameter
Number of sensors
Placement of sensors
Range
Accuracy
UV monitors
Item
Number of monitors
Wavelength range
Absorbance range
ÄKTA avant User Manual 29035184 AE
Data
7, expandable to 14
7, expandable to 14
7, expandable to 14
4
Data
4
The System pressure monitor is located after the System
pumps.
The Pre-column pressure monitor and the Post-column
pressure monitor are integrated in the Column valve V9-
C or V9H-C.
The Sample pressure monitor is located after the Sample
pump.
ÄKTA avant 25: 0 to 20 MPa (0 to 200 bar)
ÄKTA avant 150: 0 to 5 MPa (0 to 50 bar)
ÄKTA avant 25: ± 0.02 MPa or ± 2% whichever is greater
ÄKTA avant 150: ± 0.015 MPa or ± 1.5% whichever is greater
Description
Up to 2
U9-M: 190 to 700 nm in steps of 1 nm, up to 3 wavelengths
U9-L: 280 nm
-6 to 6 AU
9 Reference information
9.2 Module specifications